Hierarchy of error messages
In order to get the normal operation status Green, the following errors have to be resolved:
•Serious errors (Serious errors are errors that prevent the regular operation of the device)
•Probe & Sensor errors (Probe errors are errors of the connected probe that must be resolved before the device can resume regular operation)
•Power errors and scaling errors (Power errors and or out of scaling errors)
•Output Monitoring
•Simulator and Alarm mode
•Normal Operation
In case of a serious error, please disconnect the power supply and re power the device again.
If the failure still exist you may get in contact with Support department.
For example:
If an RMS-TCD-S-001 is mounted on the transmitter, and the Output 2 is not used, it is necessary to change the Output2 to unused within the HygroSoft software, otherwise
the device will show you a constant error mode.
To prevent an error message within the current Output Monitoring, the current version should have at least a 100 Ohm Resistor applied to the OUT1- and OUT2-.
See more:
•Serious and Probe errors: Probe Error Messages
•Power and scaling errors: Operation Error Messages
•Simulator and Alarm Mode: Simulator and Alarm
•Output Monitoring errors: Output Monitoring
•Detailed LED Status: LED status light
•Normal Operation