Function  Overview

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Transmitter > HF5A > Customer Adjustment >

Function  Overview

The device specific functions are described in the chapter Devices: HF5A

 

Hierarchy of error messages

 

In order to get the normal operation status Green, the following errors have to be resolved:

 

Serious errors (Serious errors are errors that prevent the regular operation of the device)

Probe & Sensor errors (Probe errors are errors of the connected probe that must be resolved before the device can resume regular operation)

Power errors and scaling errors (Power errors and or out of scaling errors)

Output Monitoring

Simulator and Alarm mode

Normal Operation

 

In case of a serious error, please disconnect the power supply and re power the device again.

If the failure still exist you may get in contact with Support department.

 

 

For example:

If an RMS-TCD-S-001 is mounted on the transmitter, and the Output 2 is not used, it is necessary to change the Output2 to unused within the HygroSoft software, otherwise

the device will show you a constant error mode.

 

To prevent an error message within the current Output Monitoring, the current version should have at least a 100 Ohm Resistor applied to the OUT1- and OUT2-.

 

See more:

 

Serious and Probe errors: Probe Error Messages

Power and scaling errors: Operation Error Messages

Simulator and Alarm Mode: Simulator and Alarm

Output Monitoring errors: Output Monitoring

Detailed LED Status: LED status light

Normal Operation