Simulator Mode
Used to make the HF5A generate fixed values for the humidity, temperature and calculated parameter. Can be configured, enabled or disabled.
Simulation values are a value set by the user for a specific measured value. Simulator values may already have been activated in the sensor.
Simulations values that are activated within the HF5A has priority over the probe simulation value.
Fixed simulation values from the Transmitter HF5A are marked with: Value fixed.
Fixed simulation values from Probe are marked with : (Probe:) Value fixed
Fixed simulation value of Transmitter HF5A and fixed value from Probe:
Alarm Mode
The following alarm thresholds exist for the various measurements:
• Low Alarm: Lower alarm threshold, below which an alarm is shown on the optional display.
• High Alarm: Upper alarm threshold, above which an alarm is shown on the optional display.
• Hysteresis: If a set alarm threshold is exceeded or undercut, the alarm is activated. The activated alarm is only reset if the current measured value falls
below the exceeded alarm threshold by the value of the set hysteresis or exceeds the alarm threshold by the value of the set hysteresis.
To resolve the Power Error connect OUT1+ and OUT2+, respective OUT1- and OUT2-, or change the Analog Output 1, respective Analog Output 2 to Unused,
with the HygroSoft software, when not connected with power supply.