The PCD probe series is developed for use with all RMS data loggers. Compatible from firmware version RMS-LOG-868/915 (≥V1.5), RMS-LOG-L (≥V1.4) and the RMS-LOG-L-D.

 

The PCD can be operated independently using the RTU Modbus protocol, for example for integration into OEM applications. The following functions are available via the digital interface:

oRead out serial number.

oRead out measurement value.

 

Important:

When using the PCD's to measure differential pressure and the reference 0 point is daisy chained via multiple PCD's, then please only use the Variant 2 - Membrane sensors and not the Variant 1 - Thermal mass flow flow sensors. The flow sensor leaks per design and as of such is not designed for this application.

 

RMS-Config

This SW tool is used for configuration, customer adjustment and measurement using the probe out of the RMS SW.

The following information and function are available:

oProbe information: serial number, probe address, FW version and FW update

oRead out measurements

oProbe settings: Filter and simulator function

oAdjustment

 

Hardware and software compatibility

The PCD is compatible only with RMS-generation devices. To change the PCD device configuration, the RMS-CONFIG software (≥V1.1.0) is required.

 

PCD_product image

Device Manual

Unpacking a PCD-S

Variants

Dimensions

Device description

Housing

Basic initialization

Customer adjustment

Maintenance

Modbus

Technical specification

Configuration with RMS-CONFIG

Accessories

 

 

 

 

 

RMS

RMS Cloud Login

Short user manual

Downloads

Support Contact

Forum

 

PCD-S