<< Click to Display Table of Contents >> Navigation: Hardware Overview > Sensors and Probes > PCD-S > Device description |
Functions |
Description |
Measured parameters |
The PCD measures differential pressure |
Pressure measurement range |
±25/50/100/250/500 Pa |
Temperature measurement range |
-20…80 °C |
Humidity measurement range |
0…95 %rh (non-condensing) |
Diameter |
32 mm |
Length |
87 mm |
Housing material |
Polycarbonate, anthracite |
Weight |
60 g |
Power Supply |
The clip is created to be used directly connected to and powered by a RMS LAN or RF Logger. Alternatively an AC3001 / AC3001-XD (3.3 V or 5.5 V) can be used to connect the probe to a PC. |
Digital interface |
The PCD probe has an UART interface for communication using the RTU Modbus protocol. The logic level is 2.5 V. |
Cable length |
Max 5 m |
Configurable settings |
Factory default |
Unit of measurement (metric/English) |
Pa |
Psychrometric calculation |
None |
Communication protocol |
MODBUS RTU |
MODBUS address |
1 |
Functions |
Factory default |
Differential pressure adjustment |
Factory adjusted: 5 points -1x; -0.5x; 0, +0.5x; +1x measurement range). |
Differential pressure measurement |
1 s |
Differential pressure filter |
Filter enabled at 16 |
Differential pressure simulator |
Disabled |