Depending on the application, the filter should be checked regularly for soiling, and replaced if necessary.
CCD sensors have high long term stability and do not require any additional calibration as supplied. The calibration period will depend on the activation of the automatic calibration function. If activated, then the customer should define their own calibration interval. If deactivated, then Rotronic recommends a calibration every 12 months. Calibration or adjustment can be carried out using the RMS-CONFIG or RMS-CLD/WEB software.
Check on www.rotronic.com for the latest version of the firmware. In order to update the firmware, the RMS-CONFIG Software and the AC3001 cable are required. The digital probes can only be updated via RMS-CONFIG. Carry out a Firmware Update here.