Integration of the Lighthouse Apex R5 data logger into RMS

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Converter Devices > Integration of third party products into RMS  >

Integration of the Lighthouse Apex R5 data logger into RMS

Device description

The Lighthouse Apex R5 is a particle counter that measures particle concentration with two channels and two measurement ranges.

rms-converter_Lighthouse Apex R5

Important:

Please consult the Apex R5 user manual for further details.

 

 

Network configuration of the device

The Lighthouse Apex R5 is to be connected via Ethernet connection to the network.

 

To add the Lighthouse Apex R5 into the RMS, it is necessary to know the individual network configuration of the device

oIP address (fix or DHCP)

oHost name is not supported for this device

 

Note:

A software and a separate service cable are required for network configuration. Be aware that a host name is not provided. DHCP would only work permanently if the IP address would never be changed by the DHCP server. Rotronic recommends strongly to use the fixed IP address for the device as well as for the RMS-Converter.

 

 

Integration into RMS via the RMS-CONVERTER-100

Step 1

Log in to the RMS software. Select "Tools" > "Setup" > "Devices". Select the RMS-CONVERTER-100 and click on "Add/Search" devices:

 

rms-converter_Integration into RMS via the RMS-CONVERTER

Step 2

Select the device type in the drop down menu.

 

It is possible to type the IP address or host name into the "IP address" field.

 

IMPORTANT:

If the field stays empty, RMS will not search automatically.

rms-converter_lighthouse Apex R5_Integration into RMS via the RMS-CONVERTER

Step 3

The device appears: configure accordingly.

rms-converter_lighthouse Apex R5_Integration into RMS via the RMS-CONVERTER 2

 

NOTE:

The interval cannot be shorter than the interval of the RMS-CONVERTER-100.

 

IMPORTANT:

If all check marks for measurement points are not set during initial installation and you wish to expand it later, then the same device needs to be added again with the same identification and IP address (there is no need to delete the device before adding it). The desired check marks can then be set. After completing this process, the added or updated measurement points will appear.

 

NOTE:

If settings are changed with another software further to the initial installation into RMS, then RMS will not note the change. However, such a change would cause RMS to not work correctly.

Step 4

The device is added.

RMS-Converter_Confirmation of the added device

 

 

Functionality within RMS

Step 1

Device settings: log in to the RMS software. Select "Tools" > "Setup" > "Devices". Select the device.

 

rms-converter_Lighthouse Apex R5_functionality within RMS

oStatus: the device can be disabled/enabled. When disabled, the RMS-CONVERTER-100 will not send any requests.

oSettings: Any settings can be changed and confirmed by clicking on "Save".

oProperties: Show. The device properties cannot be displayed by RMS since this function is not provided. Some device properties can be displayed using the web browser. Please type the IP address into the address bar of the web browser and it appears:

 

rms-converter_Lighthouse Apex R5_functionality within RMS 5
rms-converter_Lighthouse Apex R5_functionality within RMS 2

 

IMPORTANT:

Under "Options", "the firmware update", "import firmware file", "import device definition" and "device inventory" are not supported for this device.

Step 2

Measurement point settings: log in to the RMS software. Select "Tools" > "Setup" > "Measuring point".

 

Select the measuring point:

oStatus: the measuring point can be disabled/enabled.

oSettings: Any settings can be changed and confirmed by clicking on "Save".

 

rms-converter_Lighthouse Apex R5_functionality within RMS 3

 

NOTE:

If the measurement values are out of range RMS displays a sensor error!

Step 3

Adjustment: an adjustment of the Lighthouse Apex R5 is not possible.

Step 4

Data logging: The Lighthouse Apex R5 has no device memory supported by RMS.

 

The Lighthouse Apex R5 has no internal memory, the Lighthouse Apex R5 is not battery powered. So the Lighthouse Apex R5 cannot log during a power interruption.

 

In case of a communication interruption to the RMS Server, the RMS-CONVERTER-100 logs the data of the Lighthouse Apex R5. After the interruption, RMS requests the data from the RMS-CONVERTER-100.

 

In case of a communication interruption between the RMS-CONVERTER-100 and the Lighthouse Apex R5, the Lighthouse Apex R5 would not log the data.

Step 5

Sensor error: If the particle concentration of the air is out of range of the device, RMS displays a Sensor error.

 

rms-converter_Lighthouse Apex R5_functionality within RMS 4