<< Click to Display Table of Contents >> Navigation: Hardware Overview > LOG Dataloggers > RMS-LOG-L > Exchange Digital Probe |
The logger is compatible with the following digital probes:
When exchanging a digital probe from one type to another, the logger will show a sensor error. Follow these steps to ensure that the logger reads the digital probe correctly:
1.Archive the existing measuring point(s).
2.Remove the digital probe.
3.Reset the logger (push on the button for a minimum of 8s).
4.Remove all power suppliers (PoE, 24V and batteries).
5.Wait for 5s.
6.Add the required power supplies (PoE, 24V and batteries).
7.Add the new digital probe.
8.Add the new measuring point(s).
The sensor error should have disappeared and the new measurement value should be visible within RMS.
Updated: 12.02.2025.