Operation

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Generator > S904  >

Operation

Installing Relative Humidity Instruments for Calibration

 

Relative humidity probes can easily be installed into the humidity chamber through

the ports in the door. The amount and size of the ports are supplied per customer

specification.

 

S904_probe ports

 

NOTE: Ports which are not being used should be covered with a blind stop

to exclude temperature and relative humidity infl uences from outside the

chamber.

 

NOTE: If a door without ports (A000268) is used, the probes to be tested

must be placed inside the measurement chamber. To ensure good air fl ow do

not overfi ll the chamber with probes.

 

The port adapters can be changed using a port adapter removal tool (A000265).

 

Insert the two pegs on the adapter tool into the corresponding holes on the port

adapter and turn counter-clockwise to loosen and clockwise to tighten.

 

 

 

Start-Up

 

After installing the instruments for calibration, switch on the S904 Series by using the

ON/OFF switch on the rear panel of the instrument.

 

 

 

Manual Control of Humidity and Temperature Setting

 

The desired percentage of relative humidity and temperature (in °C) can be manually

set by using the humidity and temperature setting switches when the AUTO/MAN

switches are in the MAN position. Humidity or temperature control can be enabled or

disabled individually using the associated ON/OFF switch.

 

NOTE: Sufficient time must be allowed for the S904 Series to thermally

stabilize before monitoring the humidity and temperature readings.

 

S904_RH&Temp switches

 

 

 

 

Typical Response Times for Various Step Changes

 

Typical response times for various step changes are shown in Table 1 below. T100

represents the total time taken to reach stability at the new setpoint.

 

T63 and T95 values represent the time taken to reach 63% and 95% respectively of

the step change in relative humidity and/or temperature. (Start temperature = 23°C /

Start RH = 50%).

 

S904_response time

 

The S904 Series can also be used with relative humidity generation only or with

temperature generation only.

 

For example: If the calibration of the instruments is complete and new instruments

need to be inserted, both switches can be set to OFF. This will cause the pump to

switch off in order to save energy.

 

If only temperature calibration is needed, humidity generation can be switched off.

NOTE: In this instance, the pump will also stop.

 

The minimum chamber temperature that can be reached is about 10°C (18°F) below

ambient temperature. The maximum chamber temperature is limited to about +52°C

(+125°F).

 

 

 

Maximum Humidity Levels That can be Generated by the S904 Series

 

S904_operating range

 

 

 

 

25 pin D-sub Connector - S904

 

These two connectors provide % RH and temperature outputs from the chamber control

probe. 15 free pins wired from the internal chamber connector to the front panel

connector can be used for any purpose.

 

S904_25pin

Free (Unused)

These pins are wired from the 25-pin connector inside the chamber to straight through

to the 25-pin connector on the front panel, and can be used for any purpose.

 

These pins have a maximum current rating of 100 mA, and a maximum voltage rating

of 50 V, which must not be exceeded.

 

Ground

This pin is connected to the ground of the internal power supply.

 

Control Probe Outputs, Temperature and %RH

These are fixed 0 to 10 V outputs from the control probe inside the chamber, ranged

from 0 to 100°C and 0-100% RH respectively.

 

External set point control

To enable external setpoint control, connect +5 V to this pin with respect to ground.

 

 

 

Automatic Control of Temperature and Humidity Set Points - S904

 

These instructions do not apply to the S904D.

 

To enable the S904 set points to be controlled using a voltage input:

 

S904_RH&Temp switches

 

1.Put the MAN/AUTO switches into the AUTO position.

2.Put the ON/OFF switches into the ON position.

3.Connect +5 V to pin 24 with respect to ground (pin 21) to enable external control of the set points.

The unit is now ready to accept external inputs.

 

External inputs:

 

• %RH set point

  Pin 23, Input: 0 to 10 V, Range: 0 to 100%

 

• Temperature set point

  Pin 10, Input: 0 to 10 V, Range: 0 to 100°C

 

NOTE: Connections to the external inputs should be made with respect to

ground on pin 21.

 

NOTE: The maximum temperature in the climate chamber that can be reached

is 52°C, and the minimum is 10°C (18°F) below the ambient temperature.

 

 

 

25 pin D-sub Connector - Rear Panel and Inside Chamber - S904D

 

These two connectors provide 6 channels for data acquisition, a +14.5 V supply, ground

connection and 9 free pins wired from the internal chamber connector to the rear panel

connector that can be used for any purpose.

 

S904_25pinD

Free (Unused)

These pins are wired from the 25-pin connector inside the chamber to straight through

to the 25-pin connector on the rear panel, and can be used for any purpose.

 

These pins have a maximum current rating of 100 mA, and a maximum voltage rating

of 50 V, which must not be exceeded.

 

Channels 1-2

These channels are connected to the built-in RH probe and are always logged by the

S904D Labview® Software.

 

Channels 3-8

These channels accept a 0 to 10 V input and can also be logged by the S904D Labview®

Software.

 

14.5 V Supply - PIN 25

This pin is connected to the internal power supply of the S904D and can be used to

provide power to probes inside the chamber.

 

NOTE: For safety purposes the power supply is fi tted with a thermal cut-out

that is connected to the rear panel 25-pin connector only. It is important that

this thermal cut-out is not bypassed, or the instrument may be damaged in

the event of a fault. Refer to Figures 21 and 22 for detailed drawings showing

how to correctly wire the power supply.

 

Ground - PIN 21

This pin is connected to the ground of the internal power supply.

 

Reserved – Do not use - PINS 10, 11, 12, 13, 23, 24

 

Warning

Applying a current or voltage to the Reserved PINS may cause

permanent, irreversible damage to the S904D electronics.

 

 

 

S904D Example Wiring Diagrams

 

The following diagram shows how to connect three %RH probes using the 25-pin D-sub

connector inside the chamber, and powering the probes using the internal S904D power

supply while logging their %RH and temperature outputs.

 

S904_wiring diagrams

 

The following diagram shows how to connect three % RH probes using the 25-pin D-sub

connector on the rear panel of the S904D, and powering the probes using the internal

S904D power supply while logging their % RH and temperature outputs.

 

 

S904_wiring diagrams2

 

 

S904D Data Acquisition Resolution

 

The data acquisition is made for 0 to 10 V input. The AD converter (analog to digital) is 12

bit data acquisition, so there are 4096 steps. If a device is connected with a 0 to 5 V output

the resolution goes down to 2048 steps and for an instrument with a 0 to 1 V output the

resolution goes down to 409 steps.

 

For example: If an instrument under calibration has an output of 0 to 1 V and the

temperature output range of the instrument is -40 to +60°C, (which is a total range of

100°C) then the resolution is 0.25°C. See the calculations below:

 

S904_data resolution