<< Click to Display Table of Contents >> Navigation: Humidity & Temperature Devices > Probes > HC2A-S/XX (S, SH, SM, S-HH) > Device description |
Functions |
Description |
Measured parameters |
•Relative humidity using the ROTRONIC HYGROMER® HT-1 (HC2A-S-HH: HH-1) capacitive sensor •Temperature using a PT100 1/3 Class B sensor |
Calculated parameters |
•Dew point •Frost point |
Humidity measurement range |
0…100 %rh |
Temperature measurement range |
•-50…+100 °C (-58…+212 °F) •0…+60 °C (32…+140 °F) (HC2A-S-HH) |
Accuracy |
•±0.5 %rh, ±0.1 K at 10…30 °C (HC2A-SH) •±0.8 %rh, ±0.1 K at 10…30 °C •±1.0 %rh, ±0.1 K at 10…30 °C (HC2A-S-HH) |
Diameter |
15 mm (0.6") |
Length |
•108 mm (4.25") •109 mm (4.3") (HC2A-SM) |
Housing material |
•Polycarbonate, black •Stainless steel, DIN 1.4301 (HC2A-SM) |
Standard filter |
•Polyethylene, 40 µm •Wiremesh filter, 10 μm, DIN 1.4401 (HC2A-SM) •None (HC2A-S-HH) |
Weight |
•10 g (0.35 oz) •47 g (1.65 oz) (HC2A-SM) |
Power Supply |
The HC2A requires a power supply of
Use of a voltage adapter allows powering the probe from a voltage source of 5.0...24 VDC or 5.0...16 VAC, see Cables.
|
Analog output signals |
The HC2A probe features two 0…1 V linear analog output signals. The default factory settings are as follows:
•Output 1: 0…1 V (humidity) = 0…100 %RH •Output 2: 0…1 V (temperature) = -40…60˚C |
Digital interface |
The HC2A probe has a UART interface. |
Communication protocol |
•RO-ASCII: default communication protocol used by all AirChip 3000 devices. •I2C: does not allow the networking of devices and is limited to a one way communication where HC2A automatically sends data during each refresh interval to an external device with I2C input. •Custom: this communication protocol can be used to provide compatibility of the HC2A probe with an existing communication system. The Custom communication protocol is limited to reading measurement data from the HC2A probe.
The communication protocol can only be configured by HW5 software, see HygroClip2. |
Adjustment |
•1-point or multi-point humidity calibration or adjustment •1-point or 2-point temperature calibration or adjustment •Generate a time stamp for calibrations and adjustments •Retain and view last adjustment date and adjustment values •Generate calibration and adjustment protocols |
Configurable settings |
Factory default |
Unit of measurement (metric/English) |
Metric |
Psychrometric calculation |
None |
Output 1 parameter, scale and unit |
Humidity: 0…100%RH |
Output 2 parameter, scale and unit |
Temperature: -40…60 ˚C |
Communication protocol |
RO-ASCII |
MODBUS address |
0 |
Device name |
HC2A |
Functions |
Factory default |
Humidity/temperature adjustment |
Enabling / Disabling |
Device write protection |
Disabled |
Out-of-limit value digital alarm |
Disabled |
Data recording |
Disabled |
Fail safe mode |
Disabled |
Simulator mode |
Disabled |