HH1-SK

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Sensors >

HH1-SK

HygroMer HH1-SK - The perfect Sensor for decontamination applications with Hydrogen Peroxide

 

Sterilization & decontamination - humidity sensor

 

Designed for the use in hydrogen peroxide sterilization processes and harsh environments

Sophisticated mechanical design and superior chemical resistance

Additional protection cage for condensation prevention of the Polymer itself

 

A capacitive polymer humidity sensor is a special type of humidity sensor based on the use of polymeric materials to detect changes in humidity. The sensors utilize the properties of polymers that react to moisture and thereby influence the capacitance of the sensor. The evaluation electronic convert the capacitive value into humidity value.

The HygroMer HH1-SK sensor stands for an advanced chemical resistance and can be quoted for chemical and industrial measurements.

 

Applications

 

Decontamination & sterilization processes

Sterilization of medical equipment, incubators and clean rooms

Pharmaceutical and Chemical industries, processes and manufacturing

 

 

HH1-SK Dimension

 

HH1-SK_Hygromer

Innovation in Humidity Measurement

Designed for the use in H2O2

Advanced chemical resistant sensor

Response time < 15 s

0...100 %RH

-50...100 °C

PTFE surface protection

 

 

 

 

Further Info

Support

Contact

 

 

 

 

 

Functions

Description

Hygromer_HT-1_characteristics

Characteristic polynomial

 

5th degree polynomial

Y = A0 + A1*x + A2*x2 + A3*x3 + A4*x4 + A5*x5

 

A0 = 150

A1 = 0.45064

A2 = -8.8894E-03

A3 = 1.64369E-04

A4 = -1.3683E-06

A5 = 4.2252E-09

clip0055

Specifications

 

Capacity:

150 pF ±50 pF

Humidity operating range:

0…100 %RH

Temperature operating range:

-50...120 °C

Accuracy at 23 °C:

±1.5 %RH

Hysteresis (4 hours each at

15 %RH - 90 %RH - 15 %RH):

< 1.0 %RH

Response time τ63:

<10 s (at 23 °C and 1 m/s)

Long-term stability:

<1 %RH / year

Uncompensated

temperature deviation:

-0.15 %RH/°C

(30…90 %RH)

Frequency range

(without DC components):

1...50 kHz

Max. Voltage:

±12 VDC

 

 

The shown data are guide values. The resistance of the sensor strongly depends on the temperature and humidity conditions as well as on

exposure duration to the pollutant. Allowed fault caused from the pollutant: ±2 %RH (10…90 %RH).

 

Pollutant

Formula

Max.Workplace Allowed Concentration

 

Allowed Concentration

Continuous Operation

(ppm)

(mg/m3)

(ppm)

(mg/m3)

Hydrogen Peroxide

H2O2

10

25

800

1200