<< Click to Display Table of Contents >> Navigation: Hardware Overview > Sensors and Probes > CCA-S-20X & CCA-S-20X-SET > Technical specifications |
Functions |
Description |
Measured parameters |
The CCA measures carbon dioxide |
Technology |
NIBR |
Range |
0...20%CO2 |
Accuracy |
0...8%CO2 ±0.15%CO2 0...20%CO2 ±0.2%CO2 |
Long term stability |
±0.24%CO2/Year |
Temperature dependence |
±10% of reading |
Pressure dependence |
±0.15% of reading/hPa |
Range of application |
-20...50°C 0...100%rh non condensing 700...1200hPa |
Startup time |
60s |
Power supply |
CCA-S-20X: 3...5 VDC CCA-S-20X-SET: 12 VDC, 0.5 A |
Output signal |
CCA-S-20X: 0...2.8 V CCA-S-20X-SET: 4...20 mA |
Enclosure material |
Polycarbonate (housing) Stainless steel DIN 1.4305 (nut) |
Fire protection class |
Corresponds to UL94-HB |
Dimensions |
CCA-S-20X: 32 mm x 87 mm Converter box: 100 mm x 77 mm x 40 mm (LxWxH) |
IP protection class |
IP40 |
Weight |
CCA-S-20X: 55 g CCA-S-20X-SET: 255 g |
Functions |
Factory default |
Sensor zero |
Factory adjusted |
Sensor span |
Factory adjusted |
Sensor calibration |
Factory calibrated |
4...20 mA loop calibration |
Factory calibrated |
Example Accuracy Overview
Reading: 5.23%CO2.
Temperature: 37°C.
Pressure: 1050 hPa.
Calibration pressure: 1000hPa.
Sensor age: new.
Measurement Accuracy:
5.23%CO2 ±0.15%CO2 = 5.08%CO2<x<5.38%CO2.
Temperature Dependence:
5.23%CO2 ±10% of reading = 4.707%CO2<x<5.753%CO2.
Pressure Dependence:
Step 1: 5.23%CO2 ±0.15% of reading/hPa = 0.007845%CO2/hPa.
Step 2: Pressure difference: 1050hPa-1000hPa=50hPa.
Step 3: 0.007845*50=0.39225%CO2.
Total: 4.83775%CO2<x<5.62225%CO2