Configure your Validation

<< Click to Display Table of Contents >>

Navigation:  Software Overview > RMS Monitoring Software > Working with the RMS Software > Configure RMS >

Configure your Validation

The RMS Validation function is an automated black box test script that can be programmed and carried out. The functionality of the complete system can be tested for correctness with a validation script. In this way, changes and extensions in the system can be validated very easily and tested again at any point in time.

 

For GxP applications, the test scripts that can be generated within RMS output an operational qualification (OQ) test report, that are derived from the functional requirement specifications (FRS).

 

Each test script will deliver a test result that in turn will generate a test report.

 

Important:

oThe measured values simulated for the validation are not recorded in the SQL database! The live view (Chart / Table / Dashboard) is updated with the simulated values. These values will be replaced with the real values once the cache is refreshed.

oAs the data is not recorded in the SQL database, the simulated values generated during the validation is not taken into account within the data analysis.

oSMS and telephone alarms are not triggered during a validation routine. Only the triggering of the events is checked.

oE-Mails are triggered during the validation, each title contains "Validation". The validation script will check the audit trail 9 times at a 1 second interval to find the finished "Send E-Mail" event. Should the E-Mail not be sent out within the 9s, the audit trail will show an error. In order to check the time required for an E-Mail to be sent out, it is possible to check the event within the audit trail (under details) or check the service manager log file.

oWhen a validation is started, the devices of the measuring points involved are switched offline for one minute for monitoring purposes. The measured values are stored locally in the data logger during this time and downloaded after the validation.

oEvents that are triggered by a validation routine are marked "Validation".

 

By creating new validation script, user shall define the type of the test (for device or for measuring point). This setting cannot be changed later.

 

The following functions are available for creating and editing validations:

Function

Description

Filter

Filter the list of validations on the basis of the entry

New

Add new validation

Save

Save changes

Options

Delete, import CSV, export CSV or run test

 

Every validation has the following properties:

Property

Description

Validation info

ID

Unique script ID

Name

Freely selectable name

Test type

Device or measuring point test

Description

Freely selectable description

Test target(s)

Device or measuring points to be tested

Test script

List of test steps with command and parameter

 

Tests

Test step 1…100

Test step, definable by the user. Every test step has the following properties:

Command

 

The following commands are available per drop down menu:

oSend measured value

Instructs the measuring point (target) to output the desired measured value (parameter).

oSet Sensor Error

Simulates the sensor error status of a measuring point.

oSet alarm delay

Set alarm delay time of a measuring point. Time in seconds.

oSet battery value

Simulates battery value.

Parameter: < 10% triggers "Low battery". Parameter: > 20% clears "Low battery"

oSet Timeout

Simulates device timeout.

Parameter: timeout in seconds

oCheck alarm status

Checks the current alarm status (parameter) of the corresponding target.

oCheck send email

Checks if an email with the corresponding alarm status (parameter) was sent for the selected target.

oCheck send SMS

Checks if an SMS with the corresponding alarm status (parameter) was sent for the selected target.1

oCheck telephone call

Checks if a telephone call with the corresponding alarm status (parameter) was made for the selected target.2

Parameter

 

Depending on the command, a measured value or the alarm status can be defined here.

Execute

It is possible to execute the script via "Tools" > "Validation"

 

Actions:

oStart

Run the script

oAbout

Information from a running test

oCheck box "Continue after error"

When selected, continue with the test even after error

oCheck box for measuring points / devices

Selection of target to be tested

oSave report

Save a PDF test report. Report is saved under "Tools" > "Documents" > "Reports". File name is same as script name.

1 An SMS is not sent; the system only checks the events list.

2 A telephone call is not made; the system only checks the events list.