Maintenance

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Sensors and Probes > HCD-Sx >

Maintenance

Filter cleaning

Depending on the application, the filter should be checked regularly for soiling, and replaced if necessary.

 

 

Regular calibration

HCD sensors have high long term stability and do not require any additional calibration as supplied.

 

Long term stability is better than 1 %rh per year. We recommend calibrating the probe every 6 to 12 months, depending on the application. Calibration or adjustment can be carried out using the RMS-CONFIG or RMS server software.

 

Firmware update

Check on www.rotronic.com for the latest version of the firmware. In order to update the firmware, the RMS-CONFIG Software and the AC3001 cable are required. The digital probes can only be updated via RMS-CONFIG. Carry out a Firmware Update here.