Mechanical installation
For best measurement results, observe the following points:
oAvoid installing the sensor in the immediate vicinity of a heating element, cooling coil, cold or hot wall, in direct sunlight.
oIf the sensor is wall mounted it must not be mounted directly above a heat generating element such as a heater (rising of warm air)
oIf possible choose a location where there is good air movement at the sensor.
oAn air speed of at least 1m/s is optimal.
Use Rotronic extension cables. These allow a maximum length of 5 m.
Connector
All TCD probes use the same connector (male).
Connection diagram (7-pin connector, male – view: probe side)
|