Customer Adjustment

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Sensors and Probes > TCD-S > Maintenance >

Customer Adjustment

Temperature: the temperature signal of the device is adjusted based on the calibration points present in memory. The type of adjustment depends on the number of calibration points:

1 calibration point: offset adjustment (equivalent to a 1-point adjustment).

2 calibration points: offset and slope adjustment.

Adaptation of the A, B and C coefficients.

 

Adjustment accuracy depends both on the number and on the distribution of the calibration points over the temperature range to be measured.

 

IMPORTANT:

Be sure that the reference environment is stable and that the instrument to be adjusted has equilibrated with the environment.

Always verify that the instrument temperature measurement is accurate.