<< Click to Display Table of Contents >> Navigation: Hardware Overview > Sensors and Probes > TCD-S > Maintenance > Customer Adjustment |
Temperature: the temperature signal of the device is adjusted based on the calibration points present in memory. The type of adjustment depends on the number of calibration points:
•1 calibration point: offset adjustment (equivalent to a 1-point adjustment).
•2 calibration points: offset and slope adjustment.
•Adaptation of the A, B and C coefficients.
Adjustment accuracy depends both on the number and on the distribution of the calibration points over the temperature range to be measured.
IMPORTANT:
•Be sure that the reference environment is stable and that the instrument to be adjusted has equilibrated with the environment.
•Always verify that the instrument temperature measurement is accurate.