Accessories

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Sensors and Probes > HCD-Sx >

Accessories

Filters and Filter carriers

For reason of a sensor protection various kind of filters and filter carriers can be used. With the use of different filters, some measurement characteristics vary as well.

 

 

Filter carrier

Order code

Image

Drawing

Material

Temperature range

SPA-PCB

HCD-Sx_filtercarrier_SPA-PCB
HCD-Sx_filtercarrier_SPA-xx_drawing

Polycarbonate, black

-50…+100 °C

(-58…+212 °F)

 

SPA-PCW

HCD-Sx_filtercarrier_SPA-PCW
HCD-Sx_filtercarrier_SPA-xx_drawing

Polycarbonate, white

-50…+100 °C

(-58…+212 °F)

 

SPA-SS

HCD-Sx_filtercarrier_SPA-SS
HCD-Sx_filtercarrier_SPA-xx_drawing

Stainless steel, 1.4301

-100…+200 °C

(-148…+392 °F)

 

 

 

Filter


Only carrier

Wire mesh filter

PE filter

PTFE filter

Stainless steel

Filter

HCD-Sx_onlycarrier

HCD-Sx_wiremeshfilter

HCD-Sx_PE_PTFEfilter

HCD-Sx_PE_PTFEfilter

HCD-Sx_stainlesssteelfilter

Material

PC / 1.4301

1.4401

Polyethylene

PTFE

1.4404

Pore size

-

10 µm

40 µm

10 µm

25 µm

Temperature range

-

-100…+200 °C

-50…+100 °C

-80…+200 °C

-100…+200 °C

Response time %rh1

12 s

12 s

15 s

18 s

15 s

Response time °C 2

80 s

180 s

180 s

170 s

-

Response time °C 3

120 s

190 s

210 s

210 s

200 s

Waterproof

(immersion 50 mm)

No

No

Yes

Yes

No

Maximum wind speed

5 m/s

50 m/s

50 m/s

50 m/s

70 m/s

Suitable for particulates

No

No

>10 um

Yes

No

General information

Fast response time

High temperature, fast response time, fast drying, mechanically robust

Standard filter

High chemical resistance, particulates, high temperature

High wind speed, abrasive environment, high mechanical protection

Applications

Sterilization, H2O2, handheld devices

Meteorology, agriculture, climate chambers, storerooms, cheese cellars

HVAC, offices, storerooms, labs, clean rooms

Clean rooms, medicine, spray paint booths, flour production

HVAC with high wind speed, industrial applications

1 T63: 100 %rh => 30 %rh

2 PC carrier: 70 °C => 23 °C

3 Stainless steel carrier 70 °C => 23 °C

 

 

 

Cables

AC3001 (connect the RMS-HCD to a laptop to read out with the RMS-CONFIG software).

E2-xxA (extension cable, replace the xx with the length required: 1, 2 and 5m).

AC3003 (signal booster cable for lengths over 5m and up to 100m).

E2-05XX-MOD (UART to RS485 converter cable).