Temperature Adjustment

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Probes > PC-Series > PCMini52 > Modbus RTU Manual >

Temperature Adjustment

Select Temp Adjustment.

 

PSclip0042

 

There are 2 points which can be adjusted.

 

Point 1 is used for a low temperature adjust (zero)

Point 2 is used for a high temperature adjust (span)

 

 

 

Low temperature adjust.

 

Put the PC-mini52-ModBus RTU together with a Reference temperature meter into a stable environment with a “low” temperature and let it stabilize for a while.

 

PSclip0043

 

Select Adjust Point 1

Read the temperature of the PC-mini52-ModBus RTU at T Measured Value (5,0)

Type the value of the reference temperature meter at New Reference Value. (6,0)

Press Confirm.

 

The above temperatures can now be found in the Present Temperature Adjustments in Probe

 

Press Check new output to see the new output value

 

 

High temperature adjust.

 

Put the PC-mini52-ModBus RTU together with a Reference temperature meter into a stable environment with a “high” temperature and let it stabilize for a while.

 

PSclip0044

 

Select Adjust Point 2

 

Read the temperature of the PC-mini52-Modbus RTU at T Measured Value (59,9)

Type the value of the reference temperature meter at New Reference Value. (61,3)

Press Confirm.

 

The above temperatures can now be found in the Present Temperature Adjustments in Probe

 

Press Check new output to see the new output value

 

 

Reset to Factory Calibrations.

 

To undo the adjustments, press Reset to factory Calibration.

The next message will pop up

Click on Yes, I’m Sure to continue

 

The adjustment points will be set back to zero.

 

Press Check new output to see the new output value