<< Click to Display Table of Contents >> Navigation: Humidity & Temperature Devices > Handheld & Logger > HL-20D > Technical data |
General |
|
Device type |
Humidity-temperature data logger with internal sensor(s) |
Battery type |
3 x AA (1.5 V) alkaline |
Battery life time |
Depends on log interval and other settings. Examples based on a 30 minutes log interval: •HL20D with display sleep function set to 1 minute: 13 months |
Low battery indication |
Yes (with HygroSoft software or with optional display) |
Humidity measurement |
|
Sensor |
ROTRONIC Hygromer ® IN-1 |
Measuring range |
0…100 %RH |
Measurement accuracy at 23 °C |
±1.3 %RH @ 0…10 %RH ±0.8 %RH @ 10…60 %RH ±1.3 %RH @ 60…100 %RH |
Repeatability |
0.3 %RH |
Long term stability |
< 1 %RH / year |
Response time |
Typical 3 min. (depends on air flow) |
Temperature measurement |
|
Sensor |
Pt100 RTD, IEC 751 class A |
Measuring range |
-10...60°C |
Measurement accuracy at 23 °C |
±0.3 °C |
Repeatability |
0.05 °C |
Long term stability |
< 0.1°C / year |
Response time |
Typical 10 min. (depends on air flow) |
Calculated parameters |
|
Psychrometric Parameters |
Dew point (Dp) or Frost point (Fp) |
Startup Time and Data Refresh Rate |
|
Startup time |
1.9s (typical) |
Data refresh rate |
1.7s (typical) |
Data recording and download |
|
Recorded parameter(s) |
Relative humidity and temperature |
Recording media |
Internal EEPROM non volatile, not removable |
Recording capacity |
20,000 value pairs |
Real time clock |
Integrated, accuracy 2 s/year |
Logging modes |
Start-stop (recording stops when memory is full Loop: oldest data is dumped to make room for latest data |
Log interval |
Min. 5 sec, up to1 hour in 5 sec. increments |
Interface for data download |
Internal connector, located under protective cover |
Service and data connector |
|
Interface type |
UART (Universal Asynchronous Receiver Transmitter) |
Maximum service cable length |
5 m (16.4 ft) |
General specifications |
|
Optional display |
LC, 1 or 2 decimals resolution, backlight, alarm and low battery indication |
LED indicator |
5 sec, green flash when recording data 5 sec. red flash for out-of-limit value or logger problem |
Housing material |
ABS |
Housing protection grade |
IP40 |
Overall dimensions |
130 x 72 x 32 mm (5.1 x 2.8 x 1.3”) |
Weight |
About 210 g (7.4 oz) |
Environmental limits |
|
Operating limits |
0….50 °C 0…100 %RH, non condensing |
Critical environments |
Humidity sensor: IN-1 |