<< Click to Display Table of Contents >> Navigation: Humidity & Temperature Devices > ATEX Devices > Technical Data |
General |
|
Device type |
Humidity & temperature transmitter with analog output signals |
Circuit type |
Two-wire current loop 4 ... 20 mA |
Mounting types |
D (duct) and W (wall) |
Power Supply and Connections |
|
Supply voltage (VDD) |
10 ... 28 VDC *Load resistance in ohm. |
Rated current consumption |
2 x 20 mA |
Electrical connections |
Connections: Ex-e terminals(0.2…2.5 mm2) Cable gland: 16 x 1.5 (Ø cable 4.5…7 mm) |
Polarity protection |
Protective diode on V+ |
Calculated Parameters |
|
Psychrometric Parameters |
Dew point (Dp) above and below freezing point Frost point (Fp) below freezing point and dew point above freezing point Wet bulb temperature (Tw) Enthalpy (H) Vapor concentration (Dv) Specific humidity (Q) Mixing ratio by weight (R) Vapor concentration at saturation (Dvs) Vapor partial pressure (E) Vapor saturation pressure (Ew) |
Startup Time and Data Refresh Rate |
|
Startup time |
<60 s |
Data refresh rate |
20 s - 240 s, adjustable with the HygroSoft |
Configurable Analog Outputs |
||
Output 1 |
Can be assigned to every parameter |
|
Default parameter |
Relative humidity or calculated parameter |
|
Default range |
Per order code |
|
Output 2 |
Can be assigned to every parameter |
|
Default parameter |
Temperature or calculated parameter |
|
Default range |
Per order code |
|
Output 1 and output 2 |
||
Signal type |
4…20 mA (current loop) |
|
User-configurable range limits |
-999.99 ... +9999.99 units |
|
Short circuit tolerant |
Yes |
|
Max. load resistance |
500 Ω |
|
Min. load resistance |
0 Ω |
Service Connector |
|
Interface type |
UART |
Max. length of service cable |
5 m (16.4 ft) |
General Specifications |
|
Optional display |
LC, resolution 1 or 2 decimals, no backlight, trend and alarm indicators |
Probe material |
Stainless steel (1.4301) |
Probe dust filter material |
Sintered steel filter (1.4401) |
Housing material |
Aluminum |
Housing protection grade |
IP 66 |
Dimensions |
See Models |
Weight |
800 g |
Conformity with Standards |
|
CE / EMC Conformity |
EMC-Directive 2014/30/EU, ATEX-Directive 2014/34/EU |
Solder type |
Lead free (RoHS-Directive 2015/863/EU) |
FDA / GAMP directives |
Conform (depending on the software) |
Environmental Limits |
|
Storage and transit |
-50 ... +70 °C / -20 ... +70 °C (models with display), 0 ... 100 %RH, non-condensing |
Operating limits at electronics |
-40 ... +60 °C / -20 ... +60 °C (models with display), 0 ... 100 %RH, non-condensing |
Temperature limits at probe |
Depends on probe model |
Maximum humidity at sensor |
100 %RH to 80 °C (176 °F) 75 %RH at 100 °C (212 °F) 45 %RH at 125 °C (257 °F) 15 %RH at 150 °C (302 °F) |
Maximum air velocity at probe |
20 m/s (3935 ft /min) |
Critical environments |
Humidity sensor: IN-1/WA-1 |