MP102H/402H

<< Click to Display Table of Contents >>

Navigation:  Software > HygroSoft > Supported devices >

MP102H/402H

 

Add device - Interface

This device supports USB (with a AC3006/9 accessory cable).

 

 

Configuration

To display all the possible settings for the device, it has to be read out first by clicking on the button "Read".

By clicking on the button "Write" all the changed settings will be applied to the device.

 

Hygrosoft_MP102H

 

 

Fix values

For simulation reason, fix values can be defined and activated.

Hygrosoft_MP102H1

 

 

Device

Monitor probe alarm: The probe alarm monitoring can be enabled. Alarms could be read out via the digital interface.

Description: A device name can be defined for the individual device by the user.

Calculation: Several calculation can be chosen that are based on humidity and temperature - e. g. dew point, frostpoint, vapor pressure etc.

Fix pressure value (hPa): Enter here the fixed numerical value to be used for barometric pressure when calculating the following parameters: Wet bulb temperature, Enthalpy, Specific humidity and Mixing ratio by weight. This numerical value should correspond to the typical barometric pressure at your elevation (or in your process) and should be consistent with the unit system that is being used.

Probe description: The probe device name can be defined for the individual device by the user.

Hygrosoft_MP102H2

 

 

Analog Out

Parameter and

Scaling

 

Hygrosoft_MP102H3

 

 

Language settings

Metric and imperial can be switched for the temperature.

The unit of relative humidity can be defined.

 

Hygrosoft_MP102H4

 

 

Alarm

Alarm conditions can be defined for humidity, temperature and the calculated parameter. Values that are below the low alarm value or above the high alarm value will trigger an alarm.

The value specified for the alarm function hysteresis is used for both the low and the high alarm.

 

Hygrosoft_MP102H6

 

 

Adjust

The MP102H/402H has the possibility to use the HygroClip2 to acquire and then adjust multiple adjustment points for temperature and humidity.

For the humidity it's possible to adjust up 100 measurement points.

For the temperature it's possible to adjust up to 2 measurement points.