HygroClip2

<< Click to Display Table of Contents >>

Navigation:  Software > HygroSoft > Supported devices >

HygroClip2

 

Add device - Interface

This device supports USB (with a AC3001 accessory cable).

 

Note: If the HC2A-AW-USB cannot be added, please make sure that the USB driver is installed, see here: Driver Installation

 

 

Configuration

To display all the possible settings for the HygroClip2, it has to be read out first by clicking on the button "Read".

By clicking on the button "Write" all the changed settings will be applied to the device.

Hygrosoft_DeviceConfiguration

 

Device

Calculation: Several calculation can be chosen that are based on humidity and temperature - e. g. dew point, frostpoint, vapor pressure etc.

Description: A device name can be defined for the individual device by the user.

Hygrosoft_DeviceConfiguration1

 

 

Language settings

Metric and imperial can be switched for the temperature.

The unit of relative humidity can be defined.

 

Hygrosoft_DeviceConfiguration2

 

 

Analog output settings

Parameter and

Scaling

Hygrosoft_DeviceConfiguration3

 

 

Fix values

For simulation reason, fix values can be defined and activated.

Hygrosoft_DeviceConfiguration4

 

 

 

Alarm

Alarm conditions can be defined for humidity, temperature and the calculated parameter. Values that are below the low alarm value or above the high alarm value will trigger an alarm.

The value specified for the alarm function hysteresis is used for both the low and the high alarm.

Hygrosoft_DeviceConfiguration5

 

 

Special device configuration

The device settings can be protected with a password.

The RS485 address can be adjusted manually.

Hygrosoft_DeviceConfiguration6

 

 

Adjust

The HygroClip2 has the possibility to acquire and then adjust multiple adjustment points for temperature and humidity.

For the humidity it's possible to adjust up 100 measurement points.

For the temperature it's possible to adjust up to 2 measurement points.