<< Click to Display Table of Contents >> Navigation: Humidity & Temperature Devices > Transmitter > HF4 > Customer adjustment > Function Overview |
Overview of the main software functions of the device
MEASUREMENT ACCURACY AND RELIABILITY |
|
AirChip 3000 Functions |
Description |
Humidity / temperature adjustment |
•1-point or multi-point humidity calibration or adjustment •1-point or 2-point temperature calibration or adjustment •Generate a time stamp for calibrations and adjustments •Retain and view last adjustment date and adjustment values •Generate calibration and adjustment protocols |
MEASUREMENT LOOP VALIDATION |
|
AirChip 3000 Functions |
Description |
Simulator mode |
Used to make the HF4 generate fixed values for the humidity, temperature and calculated parameter. Can be configured, enabled or disabled |
DEVICE SAFEGUARDS |
|
AirChip 3000 Functions |
Description |
Device write protection |
Used to protect the HF4 with a password to prevent unauthorized digital access by a digital user. Can be configured, enabled or disabled |
Internal menu access from keypad |
Used to prevent accidental changes to the HF4 settings and temperature-humidity adjustment by disabling the MENU key on the optional keypad. Can be enabled or disabled |
PROCESS PROTECTION / PROTECTION OF OTHER DEVICES |
|
AirChip 3000 Functions |
Description |
Limit humidity output to 100 %RH |
Used to prevent the humidity signal from exceeding 100 %RH when condensation forms on the sensor. Can be enabled or disabled |
Out-of-limit value alarm |
Used to specify the normal range for humidity, temperature and the calculated parameter depending on the user application. Can be configured, enabled or disabled
Out-of-limit values trigger a digital alarm which can be also be seen on the optional display |