Periodic calibration check of the probe

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Transmitter > HF5 > Maintenance >

Periodic calibration check of the probe

Periodic calibration check

Both the Pt 100 RTD temperature sensor used in the probe and associated electronics are very stable and should not require any calibration after the initial factory adjustment.

Long term stability of the ROTRONIC Hygromer HT-1 humidity sensor is typically better than 1 %RH per year. For maximum accuracy, calibration of the probe should be verified every 6 to 12 months. Applications where the probe is exposed to significant pollution may require more frequent verifications.

 

Note: the HygroClip2 Advanced probe cannot be adjusted when connected to model HF52 (2-wire circuit type)

 

 

 

Procedure for adjusting the probe using the optional keypad (except model HF52)

 

Temperature adjustment

 

The keypad of the HF5 allows only a 1-point adjustment of temperature. The effect of a 1-point temperature adjustment is to add the same offset to all measured values.

When temperature is stable, press the MENU key to show the internal menu on the display

Use the (-) key to select Temperature Adjust and press the ENTER key

Make sure that the text line beginning with RefValue is highlighted and press the ENTER key

Use the (+) or (-) key to change the reference value to match the temperature reference

Use the (-) key to highlight the Adjust text line and press the ENTER key

The HF5 confirms the adjustment with the message “Adjust OK”

Press the MENU key twice to exit the menu and return the HF5 to normal operation

 

Notes:

 

Any temperature adjustment should be done prior to adjusting humidity.

The calibration point is automatically deleted from the probe memory after an adjustment

Because the HF5 has no real time clock, the date of the adjustment is not written to the probe. If retaining the adjustment date is important, use the HW5 or HygroSoft software to adjust the probe.

 

Humidity adjustment

 

The keypad of the HF5 allows a multi-point adjustment of humidity. The effect of a humidity adjustment depends on the number of calibration points present in the probe memory prior to the adjustment:

 

one calibration point: offset added to all measured values

two calibration points: offset and slope

three or more calibration points: offset, slope, linearity

 

For maximum accuracy use at least 3 to 4 calibration points distributed equally across the measurement range of interest. The calibration points (maximum 100) can be acquired in any order but we recommend going from low humidity values to high humidity values.

 

When humidity is stable, press the MENU key to show the internal menu on the display

Use the (-) key to select Humidity Adjust and press the ENTER key

Make sure that the text line beginning with RefValue is highlighted and press the ENTER key

Use the (+) or (-) key to change the reference value to match the reference humidity

Use the (-) key to highlight the Acquire text line and press the ENTER key

Note that the “Acquired” counter is incremented by 1 (number of calibration points in the probe memory)

When all calibration points have been acquired, use the (-) key to highlight the Adjust text line and press the ENTER key. Do not adjust the probe before having acquired all calibration points.

The HF5 confirms the adjustment with the message “Adjust OK”

Press the MENU key twice to exit the menu and return the HF5 to normal operation

 

Notes:

 

All calibration points are automatically deleted from the probe memory after an adjustment

Instructions for using the ROTRONIC calibration devices and humidity standards are provided in document E-M-CalBasics.

Because the HF5 has no real time clock, the date of the adjustment is not written to the probe. If retaining the adjustment date is important, use the HW5 or HygroSoft software to adjust the probe.

 

 

 

Using the HygroSoft software to adjust the probe connected to a HF5 (except model HF52):