<< Click to Display Table of Contents >> Navigation: Software > HW5 (obsolete) > General HW5 feature > Water Activity > AW measurement using AW-Therm |
Please click on the tab Water Activity on the main windows.
This tab shows a sequencer of Aw measurements that can be executed. On a first this used sequencer is still empty. To add the first Aw measurement step, please click Add on the right side.
•Add: Add a Aw measurement step.
•Edit: Existing step can be modified.
•Delete: Existing step can be deleted.
The following window opens for AW Quick or AWE (please see AW measurement modes in HW5)
•Temperature setpoint: Target temperature of the Aw Therm chamber.
•Waiting time: Waiting time before the stability criteria are checked.
•Aw mode: AW-Quick or AWE can be chosen. (please see AW measurement modes in HW5)
•AWE stability: Stability criteria of the measured Aw value. The criteria is achieved if the actual value does not differ more than +/- AWE stability value from all other values during the last minute.
•Temperature stability: Stability criteria of the measured temperature. The criteria is achieved if the actual temperature value does not differ more than +/- Temperature stability value from all other values during the last minute.
Note •The temperature stability considers only the measured temperature but does not consider the temperature set point. The user of the Aw Therm should care about the deviation between the temperature set point and the measured temperature. This deviation might be small enough for often repeated measurements. It can be achieved using the waiting time.
|
•Start: Initiates the measurement.
•Cancel: Interrupts the running process.
On the right side of the window is an overview of the settings of the chosen sequence step – see the following picture
The report can be stored.
A pdf report can be generated manually.
Please type Batch number, Product and Comments if necessary and click on Create report.
A file path should be chosen afterwards and the report is generated.