Validation

<< Click to Display Table of Contents >>

Navigation:  Help > Troubleshooting > Previous Manuals > E-OM-RMS-WEB > Data Views >

Validation

The RMS can be validated at the press of a button. The menu "Tools" > "Validation lists" all validation routines available. The creation of validation routines and further detailed information are to be found in the E-SM-RMS-WEB.

 

E-OM-RMS-WEB-V1.3.1_validation

Figure 1: Completely executed validation function

 

Function

Description

Start

Execute the selected script

Abort

Abort from an executing validation

Continue after error

Execution of the validation will continue in case of error.

 

Important Boundary Conditions:

oThe measured values sent for the validation are not recorded! The live view (Chart / Table / Dashboard) is updated with the validation points, these values are replaced with the real values after 24 hours.

oSMS and telephone alarms are not triggered during the validation routine. Only the triggering of events is checked.

oWhen a validation is started, the devices of the measuring points involved are switched offline for one minute for monitoring purposes. The measured values are stored locally in the data logger during this time and downloaded after the validation.