<< Click to Display Table of Contents >> Navigation: Software Overview > RMS Monitoring Software > Server Software Installation > RMS Software V2.3.0 > Update to V2.3.0 of the RMS Software |
Important: As soon as the Rotronic Monitoring System >V2.0.0 has been uploaded, please update all RMS-GW devices with the latest firmware version: Update Firmware.
Please reach out to Rotronic for more details: rms@rotronic.ch. |
Installing Multiple Instances.
Follow these guidelines for the installation of the V2.3.0
Prerequisites:
•A previous version of RMS is already installed
•The current version is hosted in IIS on the site “Default Web Site” and named “RMS” and “wService”
•The new RMS Files are unzipped in the folder “C:\inetpub\wwwroot\RMS\21347”
•The “connection.config” file contains the correct information for the database connection
Step 1 |
Unzip the new version and copy it into a folder with the version number parallel to the currently installed version: "C://inetpub/wwwroot/rms/{version}".
|
|
Step 2 |
Login to the SQL Server as ADMIN or SA user: |
|
Step 3 |
Create a backup of the RMS database: |
|
Step 4 |
Update the database tables using the "Rotronic-RMS-Tables".Click on "open file" and double click on the folder to update.
l
The following data will be shown and click on execute. Please check before execute in which database you will run this script. The rms database should be shown in the left top field!
|
|
Step 5 |
If the Software Version is same or higher as V1.3.X, double click on the "RMS_SQL_Indices" to upgrade and execute it. Please check with SQL Version is running on the server. Please check before execute in which database you will run this script. The rms database should be shown in the left top field!
|
|
Step 6 |
Install the .Net Framework V4.7.2 or newer, double click on the file NDP472-KB4054531-WEB.exe:
When prompted click on continue to install the .NET framework or close if it is already installed: |
|
Step 7 |
Check the existing settings can be found in "web.config" in folder \wService of the current RMS Installation. (DO NOT CHANGE ANYTHING IN THIS "web.config").
The current RMS installation can be seen under C:
Edit, if required, the connection string (via Notepad) if the in "Connections.config" in folder C:\inetpub\wwwroot\RMS\20171\ServiceManager: oData Source oInitial Catalog oUser ID oPassword
|
|
Step 8 |
Test the connection to the database, by double clicking on the "TestDatabase.cmd" command:
If the following screen appears, return to step 5 and execute the "Rotronic-RMS-Tables":
The following screen should be shown if the test is successful, you can forward to step 10:
|
|
Step 9 |
|
|
Step 10 |
Open the "Update1ToV221.cmd" or "Update2toV221.cmd" file with notepad and check to ensure that the wPageName, wServiceName and ManagerInstance are all correct.
The correct details are listed below in yellow: |
|
Step 11 |
Open the "RmsServiceManager.exe.config", in the ServiceManager file with notepad and check to ensure that the port = 8091:
Open the "NLog.config", in the ServiceManager file with notepad and check to ensure that the path is set to rms_files:
Open the "web.config", in the wPage file with notepad and check to ensure that the localhost:8091:
Open the "web.config", in the wService file with notepad and check to ensure that the localhost:8091:
Open the "NLog.config", in the wService file with notepad and check to ensure that the path is set to rms_files: |
|
Step 12 |
Right click on the "Update1ToVxx.cmd" or "Update2toVxx.cmd" and run as administrator to run the update command and follow the steps in the script:
Y
Y
When prompted, enter "Y":
The following script will run: |
|
Step 13 |
Recycle Application Pool ".NET V4.5" (in "IIS Manager" > "Application Pools" > "Recycle"). |
|
Step 14 |
Test to see if the website and the webservice are running: Login to the website
Service test: ({ServerAddresse}/wService/wService3.DeviceService.svc/TestDatabase)
|
|
Step 15 |
Only if you want to add some other report Fonts, there is the instruction.
|
|
Step 16 |
Update the Documentation path in the RMS software accordingly. The update created a new folder. The software update has created a new folder. We recommend either leaving the old folder as it is or integrating everything into the new one. If background images, layouts, etc. are used in the software, they must be copied from the old folder to the new one.
|
|
Step 17 |
Once the software updated, ensure that a new license key is also added: Update your License Code. |