RMS-CONVERTER-100 Open Issues

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Gateway and Converters > RMS-CONVERTER-100 >

RMS-CONVERTER-100 Open Issues

The RMS-CONVERTER open issues are only valid as of the most actual firmware and software versions.

 

RMS Issue Number

Products Concerned

Description

Root Cause

 

Solution

N/A

RMS-CONVERTER-100 and all Rotronic devices with time displayed.

The data logger time will always show the UTC time.

RMS system requirement.

No solution available.

RMS-CONVERTER-2594

RMS-CONVERTER-100 and HygroLog and HL-NT products.

The HygroLog NT/HL-NT freezes. The user needs to remove all power supplies in order to restart the HygroLog NT/HL-NT.

HygroLog NT/HL-NT issue.

No solution available.

RMS-CONVERTER-2638

RMS-CONVERTER-100 and HygroLog and HL-NT products.

The HygroLog NT/HL-NT measuring points are frequently going into sensor error and recovering.

Currently not defined.

No solution available.

RMS-CONVERTER-2427

RMS-CONVERTER-100 and HygroLog and HL-NT products.

Peaks on RMS and not on the HygroLog NT/HL-NT.

Currently not defined.

No solution available.

RMS-CONVERTER-2355

RMS-CONVERTER-100 and HF8 products.

The switching of the relay creates an event "Probe replaced" and shows "Serial number -> 0"

Currently not defined.

No solution available.

RMS-CONVERTER-2496

RMS-CONVERTER-100 V1.4 and HygroLog and HL-NT products.

Data gaps appear when reading out digital inputs.

RMS-CONVERTER-100 and RMS software issue.

RMS-CONVERTER-V1.5 and RMS V2.2.

RMS-CONVERTER-2476

RMS-CONVERTER-100 V1.4 and HygroLog and HL-NT products.

Device & Server time synchronization is missing within the audit trail. Data gaps occur.

RMS-CONVERTER-100 and RMS software issue.

RMS-CONVERTER-V1.5 and RMS V2.2.