The RMS-CONVERTER open issues are only valid as of the most actual firmware and software versions.
RMS Issue Number |
Products Concerned |
Description |
Root Cause
|
Solution |
---|---|---|---|---|
N/A |
RMS-CONVERTER-100 and all Rotronic devices with time displayed. |
The data logger time will always show the UTC time. |
RMS system requirement. |
No solution available. |
RMS-CONVERTER-2594 |
RMS-CONVERTER-100 and HygroLog and HL-NT products. |
The HygroLog NT/HL-NT freezes. The user needs to remove all power supplies in order to restart the HygroLog NT/HL-NT. |
HygroLog NT/HL-NT issue. |
No solution available. |
RMS-CONVERTER-2638 |
RMS-CONVERTER-100 and HygroLog and HL-NT products. |
The HygroLog NT/HL-NT measuring points are frequently going into sensor error and recovering. |
Currently not defined. |
No solution available. |
RMS-CONVERTER-2427 |
RMS-CONVERTER-100 and HygroLog and HL-NT products. |
Peaks on RMS and not on the HygroLog NT/HL-NT. |
Currently not defined. |
No solution available. |
RMS-CONVERTER-2355 |
RMS-CONVERTER-100 and HF8 products. |
The switching of the relay creates an event "Probe replaced" and shows "Serial number -> 0" |
Currently not defined. |
No solution available. |
RMS-CONVERTER-2496 |
RMS-CONVERTER-100 V1.4 and HygroLog and HL-NT products. |
Data gaps appear when reading out digital inputs. |
RMS-CONVERTER-100 and RMS software issue. |
RMS-CONVERTER-V1.5 and RMS V2.2. |
RMS-CONVERTER-2476 |
RMS-CONVERTER-100 V1.4 and HygroLog and HL-NT products. |
Device & Server time synchronization is missing within the audit trail. Data gaps occur. |
RMS-CONVERTER-100 and RMS software issue. |
RMS-CONVERTER-V1.5 and RMS V2.2. |