Device description

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Sensors and Probes > HCD-IC >

Device description

General information

Functions

Description

Measured parameters

Relative humidity and temperature using the ROTRONIC HYGROMER® HT-1 and a Pt100 temperature sensor

Humidity measurement range

0…100 %rh

Temperature measurement range

Electronic: -40…+85 °C

Sensor head: -100…+200 °C1

Continuous Load: 190 °C max

Accuracy

±0.8 %rh, ±0.1 K at 10…30 °C

Diameter

15 mm (0.6")

Length

Probe head: 144 mm

Cable length: 2 m (HCD-IC102)

Electronics and connector: 111 mm

Housing material

PPS (black), stainless steel 1.4301

Standard filter

non

Weight

450 g

Power Supply

The HCD-IC requires a power supply of 3.3…15 VDC

Digital interface

The HCD-IC probe has a UART interface with an logical level of 3.3 V

Modbus

Measurement values and serial number can be read out if the probe is to be used out of the RMS - please see Modbus

Adjustment

1-point or multi-point humidity adjustment/calibration

1-point or 2-point temperature adjustment/calibration

Output the last adjustment/calibration, with date

Produce an adjustment/calibration log

Sensor alarm

The probe outputs a digital alarm if the sensor has a short circuit or open circuit. This function cannot be deactivated

Cable length

maximum 3 m

1The peak load at 200 °C is 100 h maximum. Continuous Load 190 °C maximum.

 

 

Factory defaults

Configurable settings

Factory default

Unit of measurement (metric/English)

Metric

Psychrometric calculation

None

Communication protocol

MODBUS

MODBUS address

0

Device name

HCD-IC102

 

Functions

Factory default

Humidity/temperature adjustment

Factory adjusted

Write-protection for device settings

Disabled

Simulator mode

Disabled