<< Click to Display Table of Contents >> Navigation: Hardware Overview > Sensors and Probes > HCD-IC > Device description |
Functions |
Description |
Measured parameters |
Relative humidity and temperature using the ROTRONIC HYGROMER® HT-1 and a Pt100 temperature sensor |
Humidity measurement range |
0…100 %rh |
Temperature measurement range |
Electronic: -40…+85 °C Sensor head: -100…+200 °C1 Continuous Load: 190 °C max |
Accuracy |
±0.8 %rh, ±0.1 K at 10…30 °C |
Diameter |
15 mm (0.6") |
Length |
Probe head: 144 mm Cable length: 2 m (HCD-IC102) Electronics and connector: 111 mm |
Housing material |
PPS (black), stainless steel 1.4301 |
Standard filter |
non |
Weight |
450 g |
Power Supply |
The HCD-IC requires a power supply of 3.3…15 VDC |
Digital interface |
The HCD-IC probe has a UART interface with an logical level of 3.3 V |
Modbus |
Measurement values and serial number can be read out if the probe is to be used out of the RMS - please see Modbus |
Adjustment |
•1-point or multi-point humidity adjustment/calibration •1-point or 2-point temperature adjustment/calibration •Output the last adjustment/calibration, with date •Produce an adjustment/calibration log |
Sensor alarm |
The probe outputs a digital alarm if the sensor has a short circuit or open circuit. This function cannot be deactivated |
Cable length |
maximum 3 m |
1The peak load at 200 °C is 100 h maximum. Continuous Load 190 °C maximum.
Configurable settings |
Factory default |
Unit of measurement (metric/English) |
Metric |
Psychrometric calculation |
None |
Communication protocol |
MODBUS |
MODBUS address |
0 |
Device name |
HCD-IC102 |
Functions |
Factory default |
Humidity/temperature adjustment |
Factory adjusted |
Write-protection for device settings |
Disabled |
Simulator mode |
Disabled |