<< Click to Display Table of Contents >> Navigation: Hardware Overview > Sensors and Probes > HCD-Sx > Device description |
Functions |
Description |
Measured parameters |
Relative humidity and temperature using the ROTRONIC HYGROMER® HT-1 and a Pt1000 temperature sensor |
Humidity measurement range |
0…100 %rh |
Temperature measurement range |
-40…+85 °C (-40…+185 °F) |
Accuracy |
±0.8 %rh, ±0.1 K at 10…30 °C |
Diameter |
15 mm (0.6") |
Length |
108 mm (3.3") |
Housing material |
Polycarbonate, black |
Standard filter |
Polyethylene, 40 µm |
Weight |
10 g (0.35 oz) |
Power Supply |
The HCD requires a power supply of |
Digital interface |
The HCD probe has a UART interface with an logical level of 2.5 V |
Modbus |
Measurement values and serial number can be read out if the probe is to be used out of the RMS - please see Modbus |
Adjustment |
•1-point or multi-point humidity adjustment/calibration •1-point or 2-point temperature adjustment/calibration •Output the last adjustment/calibration, with date •Produce an adjustment/calibration log |
Sensor alarm |
The probe outputs a digital alarm if the sensor has a short circuit or open circuit. This function cannot be deactivated. |
Cable length |
maximum 3 m |
Configurable settings |
Factory default |
Unit of measurement (metric/English) |
Metric |
Psychrometric calculation |
None |
Communication protocol |
MODBUS |
MODBUS address |
0 |
Device name |
HCD |
Functions |
Factory default |
Humidity/temperature adjustment |
Factory adjusted |
Write-protection for device settings |
Disabled |
Simulator mode |
Disabled |