Device description

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Sensors and Probes > HCD-Sx >

Device description

General information

Functions

Description

Measured parameters

Relative humidity and temperature using the ROTRONIC HYGROMER® HT-1 and a Pt1000 temperature sensor

Humidity measurement range

0…100 %rh

Temperature measurement range

-40…+85 °C (-40…+185 °F)

Accuracy

±0.8 %rh, ±0.1 K at 10…30 °C

Diameter

15 mm (0.6")

Length

108 mm (3.3")

Housing material

Polycarbonate, black

Standard filter

Polyethylene, 40 µm

Weight

10 g (0.35 oz)

Power Supply

The HCD requires a power supply of
2.8…5.5 VDC

Digital interface

The HCD probe has a UART interface with an logical level of 2.5 V

Modbus

Measurement values and serial number can be read out if the probe is to be used out of the RMS - please see Modbus

Adjustment

1-point or multi-point humidity adjustment/calibration

1-point or 2-point temperature adjustment/calibration

Output the last adjustment/calibration, with date

Produce an adjustment/calibration log

Sensor alarm

The probe outputs a digital alarm if the sensor has a short circuit or open circuit. This function cannot be deactivated.

Cable length

maximum 3 m

 

 

 

Factory defaults

Configurable settings

Factory default

Unit of measurement (metric/English)

Metric

Psychrometric calculation

None

Communication protocol

MODBUS

MODBUS address

0

Device name

HCD

 

Functions

Factory default

Humidity/temperature adjustment

Factory adjusted

Write-protection for device settings

Disabled

Simulator mode

Disabled