<< Click to Display Table of Contents >> Navigation: Hardware Overview > Sensors and Probes > HCD-IC > Humidity and Temperature Calibration/Adjustment > Calibration |
Calibration consists in capturing humidity and / or temperature values measured by an instrument at a number of known reference conditions without making any changes to the instrument. As-Found data is the result of a calibration.
oTemperature: the HCD can retain in memory up to 2 calibration points, each consisting of: measured temperature value and reference temperature value. The calibration points are retained in memory until erased by the user or until the device is adjusted by the user. The calibration points are read by RMS whenever the Probe Adjustment form is opened. In principle, the calibration points can be saved in any order.
oHumidity: the HCD can retain in memory up to 20 calibration points, each consisting of: measured humidity value, reference humidity value and temperature value measured at the time of calibration. The calibration points are retained in memory until erased by the user or until the device is adjusted by the user. The calibration points are read by RMS whenever the Probe Adjustment form is opened. In principle, the calibration points can be generated at any temperature and saved in any order.