Calibration Option 4

<< Click to Display Table of Contents >>

Navigation:  Service Overview > Calibration in RMS >

Calibration Option 4

This option is valid for any company wishing to calibrate an RMS LAN logger.

 

Important:

The LAN logger needs to be deactivated within the RMS software.

 

Calibration is only possible with the RMS-LOG-T30-L, as the digital probes are not seen within via the RMS-LOG-L device. For all other LAN devices please see the Calibration Option 1.

 

The RMS-ADC-L-R cannot be calibrated within RMS.CONFIG, only the scaling can be made.

 

Requires:

1 * RMS LAN logger

1 * RMS-Config software.

 

Step 1

Place the LAN logger in the calibration environment.

Ensure that the logger is connected to the LAN network.

Open the RMS-CONFIG software.

Step 2

Open the RMS-CONFIG software. Click on "Device" > "Search" > "Network devices", the list of network devices will appear:

calibration option 4_1

 

 

Step 3

Select the required device and click on "Add":

 

calibration option 4_2

Step 4

Click on the Adjust tab:

calibration option 4_3

 

Step 5

Select the required input:

calibration option 4_4

 

Step 6

Follow the internal calibration procedures.

 

Enter the reference value shown on the reference and click on acquire:

calibrate digital probe with RMS-CONFIG5

Step 7

The acquired points will be added to acquired points:

calibrate digital probe with RMS-CONFIG6

Step 8

Carry out all of the necessary calibration points.

 
The data found within the acquired points is the calibration data with the as found values.

 

Important:

This is only a calibration and not an adjustment.

 

Adjust: Click on adjust to adjust the probe. Why Adjust?

 

Delete point: Delete an acquired point if required.

 

Delete all points: Delete all points if required.

 

Restore factory adjust: Restore the factory adjustment if required.