| << Click to Display Table of Contents >> Navigation: CO2 & Pressure Devices > CP11 > Trouibleshooting | 
| Error | Message | Solution | 
|---|---|---|
| E01 | The CO2 sensor is not working | Switch the instrument off and back on again | 
| E33 | The CO2 sensor is not working | Repeat the CO2 calibration | 
| E02 | The measured value is too low | Place the device in a normal environment | 
| E03 | The measured value is too high | Place the device in a normal environment | 
| E04 | The dew point and wet bulb temperature cannot be determined because the necessary values for calculation were not measured correctly. (DP, WB) | Correct the error | 
| E11 | RH calibration error | Repeat the humidity calibration | 
| E31 | Temperature sensor or AD damaged | Send the instrument in for repair | 
| E32 | Memory IC damaged | Send the instrument in for repair | 
| E33 | RH sensor or circuit damaged | Send the instrument in for repair |