| << Click to Display Table of Contents >> Navigation: Humidity & Temperature Devices > Probes > XD > Technical Specification | 
| General | |
| Humidity sensor | ROTRONIC Hygromer ® HT-1 | 
| Temperature sensor | PT100 1/3 class B | 
| Response time sensor | t63: <15 s without filter, (temperature and humidity) | 
| Max. air velocity (m/s) | 3.5 m/s without filter, 20 m/s with polyethylen filter | 
| Operating humidity | 0…100 %rh, non-condensing | 
| Operating temperature | -40… 85 °C | 
| Accuracy @23 °C | ±0,8 %RH ±0,2 K @5...24 VDC | 
| Factory adjustment | @23°C and 10, 35, 80 %rH | 
| Long-term stability | <1 %RH / year (with clean air) | 
| Supply voltage | 5...16 VAC / 5...24 VDC (0...1 V) 12...16 VAC / 16...24 VDC (all kind of output signals) | 
| Current consumption | Approx. 7 mA @ 24 VDC, 0...10 V Approx. 12 mA @ 5 VDC, 0...1 V | 
| Protection rating | IP65 (except the sensor area) | 
| Digital communication | UART (19200 baud fixed) | 
| Protocols | RoAscii (default), I2C, Modbus (not universal) | 
| Analogue outputs | 2x 0…1/5/10 VDC | 
| Analogue outputs parameters | •Humidity (default) •Temperature (default) •Dew point (setting with HygroSoft) •Frost point (setting with HygroSoft) | 
| Analog output scaling | •Humidity (0…100 %rh = 0...1 V) •Temperature (-40…60 °C = 0…1 V) •Freely settable with HygroSoft | 
| Timing | 1st measurement after 3 s Measurement interval 1 s | 
| Accuracy analog output | @23°C < 1 mV (1 V) / < 10 mV (10 V) / <20uA | 
| Air Chip 3000 | Saves adjustment data so that probes can be interchanged without re-adjusting | 
| Compatible devices | Not comaptible with other Rotronic devices | 
| Delivery package | •Probe •Short Instruction Manual •Calibration Certificate | 
Attention: The XD-series is not compatible with other Rotronic devices, as HF5, PF4/5 and HP32