<< Click to Display Table of Contents >> Navigation: Humidity & Temperature Devices > Probes > XD > Technical Specification |
General |
|
Humidity sensor |
ROTRONIC Hygromer ® HT-1 |
Temperature sensor |
PT100 1/3 class B |
Response time sensor |
t63: <15 s without filter, (temperature and humidity) |
Max. air velocity (m/s) |
3.5 m/s without filter, 20 m/s with polyethylen filter |
Operating humidity |
0…100 %rh, non-condensing |
Operating temperature |
-40… 85 °C |
Accuracy @23 °C |
±0,8 %RH ±0,2 K @5...24 VDC |
Factory adjustment |
@23°C and 10, 35, 80 %rH |
Long-term stability |
<1 %RH / year (with clean air) |
Supply voltage |
5...16 VAC / 5...24 VDC (0...1 V) 12...16 VAC / 16...24 VDC (all kind of output signals) |
Current consumption |
Approx. 7 mA @ 24 VDC, 0...10 V Approx. 12 mA @ 5 VDC, 0...1 V |
Protection rating |
IP65 (except the sensor area) |
Digital communication |
UART (19200 baud fixed) |
Protocols |
RoAscii (default), I2C, Modbus (not universal) |
Analogue outputs |
2x 0…1/5/10 VDC |
Analogue outputs parameters |
•Humidity (default) •Temperature (default) •Dew point (setting with HygroSoft) •Frost point (setting with HygroSoft) |
Analog output scaling |
•Humidity (0…100 %rh = 0...1 V) •Temperature (-40…60 °C = 0…1 V) •Freely settable with HygroSoft |
Timing |
1st measurement after 3 s Measurement interval 1 s |
Accuracy analog output |
@23°C < 1 mV (1 V) / < 10 mV (10 V) / <20uA |
Air Chip 3000 |
Saves adjustment data so that probes can be interchanged without re-adjusting |
Compatible devices |
Not comaptible with other Rotronic devices |
Delivery package |
•Probe •Short Instruction Manual •Calibration Certificate |
Attention: The XD-series is not compatible with other Rotronic devices, as HF5, PF4/5 and HP32