<< Click to Display Table of Contents >> Navigation: Humidity & Temperature Devices > Transmitter > XB32A > Technical Specification |
XB3 |
|
General |
|
Device type |
Humidity temperature transmitter |
Circuit type |
3-wire |
Power supply and connections |
|
Supply voltage (VDD) |
15…40 VDC or 13…28 VAC |
Nominal current consumption |
< 50 mA |
Electrical connections |
Terminal block and M16 cable grip |
Polarity protection |
Protective diode on V+ |
Humidity measurement |
|
Sensor |
ROTRONIC Hygromer ® HT-1 |
Measuring range |
0…100 %RH |
Measurement accuracy at 23 °C |
±1.0 %RH |
Repeatability |
0.3 %RH |
Long term stability |
< 1 %RH / year |
Sensor time constant |
Typical 10 sec, 63% of a 35 to 80 %RH step change (1m/sec air flow at sensor) |
Temperature measurement |
|
Sensor |
Pt100 RTD, IEC 751 1/3 class B |
Measuring range |
-100…200 °C (see also environmental limits) |
Measurement accuracy at 23 °C |
±0.2 °C |
Repeatability |
0.05°C |
Long term stability |
< 0.1°C / year |
Sensor time constant |
Typical 4 sec, 63% of a step change (1m/sec air flow at sensor) |
Calculated parameters |
|
Psychrometric calculations |
Dew or frost point (user configurable option) |
Start-up time and data refresh rate |
|
Start-up time |
1.9s (typical) |
Data refresh rate |
1.7s (typical) |
Digital interface (service connector) |
|
Interface type |
UART (Universal Asynchronous Receiver Transmitter) |
Maximum service cable length |
5 m (16.4 ft) |
General specifications |
|
Probe material |
PPS |
Probe dust filter material |
Depends on filter type |
Physical dimensions |
|
Weight |
225 g (8.0 oz) |
Conformity with standards |
|
CE / EMC immunity |
EMC Directive 2014/30/EU: EN 61000-6-1; EN 61000-6-2; EN 61000-6-3; EN 61000-6-4; + A11 |
Solder type |
Lead free (RoHS directive) |
FDA / GAMP directives |
compatible |
Environmental limits |
|
Storage and transit |
-50…+70 °C 0…100 %RH, non condensing |
Operating limits at electronics |
-40…85 °C, 0…100 %RH, non condensing |
Temperature limits at probe |
-100…200 °C |
Maximum humidity at sensor |
100 %RH up to 80 °C (176 °F) 70 %RH at 100 °C (212 °F) 30 %RH at 125 °C (260 °F) 15 %RH at 150 °C (302 °F) 8 %RH at 175 °C (347 °F) 4 %RH at 200 °C (392 °F) |
Maximum air velocity at probe |
40 m/s (7,870 ft /min) – depends on dust filter type |
Critical environments |
Humidity sensor: as per DV04-14.0803.02 – Critical chemicals |
Configurable analog outputs |
||
Output 1 |
Can be made to correspond to any parameter |
|
Factory default parameter |
Relative humidity or dew / frost point |
|
Factory default scale |
As per ordering code |
|
Output 2 |
Can be made to correspond to any parameter |
|
Factory default parameter |
Temperature |
|
Factory default scale |
As per ordering code |
|
Output 1 and Output 2 |
||
Signal type |
0…20 mA 4… 20 mA 0… 1 V 0… 2 V 0… 5 V 0… 10 V (user configurable) |
|
Accuracy of analog output |
±20 μA ±2 mV (0-1V) ±5 mV (0-10V) |
|
User configurable scaling limits |
-999.99 … 9999.99 engineering units |
|
Maximum offset at bottom of signal range |
0…1V signal range: 10 mV 0…2V signal range: 20 mV 0…5V signal range: 50 mV 0…10V signal range: 100 mV 0…20mA signal range: 0.2 mA 4…20mA signal range: no offset |
|
Short circuit tolerant |
Yes |
|
Maximum external load |
500 Ω (current output) |
|
Minimum external load |
1 kΩ (0…1 V output) 2 kΩ (0…2 V output) 5 kΩ (0…5 V output) 10 kΩ (0…10 V output) 0 Ω (current output) |
The XB transmitter can be configured to calculate either the dew point or frost point based on the measurement of relative humidity and temperature. The accuracy of this conversion varies, depending on the humidity and temperature conditions as shown in the graph below:
Example: At a temperature of 20 ˚C, a dew point value of -37 ˚C is measured with an accuracy of ±1.0 ˚C or better.