Technical Specification

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Transmitter > XB32A >

Technical Specification

Specification overview


XB3

General


Device type

Humidity temperature transmitter

Circuit type

3-wire

Power supply and connections


Supply voltage (VDD)

15…40 VDC or 13…28 VAC

Nominal current consumption

< 50 mA

Electrical connections

Terminal block and M16 cable grip

Polarity protection

Protective diode on V+

Humidity measurement


Sensor

ROTRONIC Hygromer ® HT-1

Measuring range

0…100 %RH

Measurement accuracy at 23 °C

±1.0 %RH

Repeatability

0.3 %RH

Long term stability

< 1 %RH / year

Sensor time constant

Typical 10 sec, 63% of a 35 to 80 %RH step change (1m/sec air flow at sensor)

Temperature measurement


Sensor

Pt100 RTD, IEC 751 1/3 class B

Measuring range

-100…200 °C (see also environmental limits)

Measurement accuracy at 23 °C

±0.2 °C

Repeatability

0.05°C

Long term stability

< 0.1°C / year

Sensor time constant

Typical 4 sec, 63% of a step change (1m/sec air flow at sensor)

Calculated parameters


Psychrometric calculations

Dew or frost point (user configurable option)

Start-up time and data refresh rate


Start-up time

1.9s (typical)

Data refresh rate

1.7s (typical)

Digital interface (service connector)


Interface type

UART (Universal Asynchronous Receiver Transmitter)

Maximum service cable length

5 m (16.4 ft)

General specifications


Probe material

PPS

Probe dust filter material

Depends on filter type

Physical dimensions

Dimensions

Weight

225 g (8.0 oz)

Conformity with standards


CE / EMC immunity

EMC Directive 2014/30/EU:

EN 61000-6-1;

EN 61000-6-2;

EN 61000-6-3;

EN 61000-6-4; + A11

Solder type

Lead free (RoHS directive)

FDA / GAMP directives

compatible

Environmental limits


Storage and transit

-50…+70 °C 0…100 %RH, non condensing

Operating limits at electronics

-40…85 °C, 0…100 %RH, non condensing

Temperature limits at probe

-100…200 °C

Maximum humidity at sensor

100 %RH up to 80 °C (176 °F)

70 %RH at 100 °C (212 °F)

30 %RH at 125 °C (260 °F)

15 %RH at 150 °C (302 °F)

8 %RH at 175 °C (347 °F)

4 %RH at 200 °C (392 °F)

Maximum air velocity at probe

40 m/s (7,870 ft /min) – depends on dust filter type

Critical environments

Humidity sensor: as per DV04-14.0803.02 – Critical chemicals

Configurable analog outputs


Output 1

Can be made to correspond to any parameter


Factory default parameter

Relative humidity or dew / frost point


Factory default scale

As per ordering code

Output 2

Can be made to correspond to any parameter


Factory default parameter

Temperature


Factory default scale

As per ordering code

Output 1 and Output 2



Signal type

0…20 mA

4… 20 mA

0… 1 V

0… 2 V

0… 5 V

0… 10 V

(user configurable)


Accuracy of analog output

±20 μA

±2 mV (0-1V)

±5 mV (0-10V)


User configurable scaling limits

-999.99 … 9999.99 engineering units


Maximum offset at bottom of signal range

0…1V signal range:       10 mV

0…2V signal range:       20 mV

0…5V signal range:       50 mV

0…10V signal range:     100 mV

0…20mA signal range:  0.2 mA

4…20mA signal range:  no offset


Short circuit tolerant

Yes


Maximum external load

500 Ω (current output)


Minimum external load

1 kΩ (0…1 V output)

2 kΩ (0…2 V output)

5 kΩ (0…5 V output)

10 kΩ (0…10 V output)

0 Ω (current output)

 

 

 

Dew point accuracy

The XB transmitter can be configured to calculate either the dew point or frost point based on the measurement of relative humidity and temperature. The accuracy of this conversion varies, depending on the humidity and temperature conditions as shown in the graph below:

 

XB-dew-point-accuracy

 

Example: At a temperature of 20 ˚C, a dew point value of -37 ˚C is measured with an accuracy of ±1.0 ˚C or better.