<< Click to Display Table of Contents >> Navigation: Humidity & Temperature Devices > Transmitter > HF5 > Technical Specification |
HF52 |
HF53 / HF54 / HF56 |
|
General |
||
Device type |
Humidity temperature transmitter with analog output signals |
Humidity temperature transmitter with analog output signals and optional digital output |
Circuit type |
2-wire, loop powered |
3-wire |
Mechanical configuration types |
D and W |
|
Power supply and connections |
||
Supply voltage (VDD) |
10…28VDC V min = 10 V + (0.02 x Load*) *Load resistance in Ω. |
HF53: 18…40 VDC or 12…28 VAC
HF54: 9 to 36 VDC or 7 to 24 VAC Galvanic isolation
HF56: 100 to 240 VAC 47 to 63 Hz Galvanic isolation |
Typical current consumption |
2 x 20 mA |
HF53 / HF54: 100 mA
HF56: < 50 mA at 100 VAC < 35 mA at 240 VAC |
Electrical connections |
Terminal blocks and 1 x M16 cable grip or ½” conduit adapter |
HF53 / HF54: Terminal blocks and 1 x M16 cable grip or ½” conduit adapter
HF56: Terminal blocks and 2 x M16 cable grip |
Polarity protection |
Protective diode on V+ (except HF56) |
|
Fuse |
HF56: 2A fuse, soldered to PCB |
|
Calculated parameters |
||
Psychrometric calculations |
Dew point (Dp) above and below freezing Frost point (Fp) below freezing and dew point above freezing Wet bulb temperature (Tw) Enthalpy (H) Vapor concentration (Dv) Specific humidity (Q) Mixing ratio by weight (R) Vapor concentration at saturation (Dvs) Vapor partial pressure (E) Vapor saturation pressure (Ew) |
|
Start-up time and data refresh rate |
||
Start-up time |
60 s (typical) |
3.s (typical) |
Data refresh rate |
20 s (typical) |
1 s (typical) |
Digital interface |
||
Interface type |
N/A |
HF53 / HF54: RS-485 or USB + RS-485 or Ethernet + RS-485
HF56: RS-485 + Ethernet |
Service connector |
||
Interface type |
UART (Universal Asynchronous Receiver Transmitter) |
|
Maximum service cable length |
5 m (16.4 ft) |
|
General specifications |
||
Optional display |
LC, 1 or 2 decimals resolution, no backlight, trend and alarm indication |
LC, 1 or 2 decimals resolution, backlight, trend and alarm indication |
Housing material |
ABS |
|
Housing protection grade |
IP 65 |
IP 65* (except models with Ethernet, only IP20) |
*Only IP65 when the red service plug is mounted |
||
Physical dimensions |
See chapter Dimensions |
|
Weight |
250 g (8.8 oz) |
Models with small enclosure: 250 g (8.8 oz) Models with large enclosure: 540 g (1 lb 3 oz) Models in aluminum enclosure: 750 g (26.5 oz) |
Impact stress |
IK08 according to IEC 62262 (horizontal impact) |
|
Degree of pollution |
2 |
|
Overvoltage category |
- |
|
Conformity with standards |
||
CE / EMC immunity |
EMC Directive 2014/30/EU: EN 61000-6-1: 2007 EN 61000-6-2: 2005 EN 61000-6-3: 2007 EN 61000-6-4: 2007 + A11: 2011 EN 50581: 2012 EN 61326-1: 2013 |
EMC Directive 2014/30/EU: HF53 / HF54: EN 61000-6-1: 2007 EN 61000-6-2: 2005 EN 61000-6-3: 2007 EN 61000-6-4: 2007 + A11: 2011 EN 50581: 2012 HF56: EN61326-1:2013 |
Solder type |
Lead free (RoHS directive) |
|
Fire protection class |
Corresponds to UL94-HB |
|
FDA / GAMP directives |
compatible |
|
Environmental limits |
||
Storage and transit |
-50…+70 °C / -20...+70 °C (models with display), 0…100 %RH, non-condensing |
|
Operating limits at electronics |
-40…+60 °C -10…+60 °C (models with display) 0…100 %RH, noncondensing |
HF53 / HF54 : -40…+60 °C -10…+60 °C (models with display)
HF56: -10…50 °C (models with/out display) 0…100 %RH, noncondensing |
Temperature limits at probe |
Depends on probe model |
|
Maximum humidity at sensor |
100 %RH up to 80 °C (176 °F) 75 %RH at 100 °C (212 °F) 45 %RH at 125 °C (260 °F) 15 %RH at 150 °C (302 °F) |
|
Maximum air velocity at probe |
20 m/s (3,935 ft /min) |
|
Critical environments |
Humidity sensor: as per DV04-14.0803.02 – Critical chemicals |
|
Maximal operating altitude |
2000 metres above sea level |
Configurable analog outputs |
HF52 |
HF53 / HF54 / HF56 |
|
Output 1 |
Can be made to correspond to any parameter |
||
Factory default parameter |
Relative humidity or dew / frost point |
||
Factory default scale |
As per ordering code |
||
Output 2 |
Can be made to correspond to any parameter |
||
Factory default parameter |
Temperature |
||
Factory default scale |
As per ordering code |
||
Output 1 and Output 2 |
|||
Signal type |
4…20 mA |
0…20 mA 4… 20 mA 0… 1 V 0… 5 V 0… 10 V (user configurable) |
|
Accuracy of analog output |
±30 μA |
±20 μA ±2 mV (0-1V) ±5 mV (0-10V) |
|
User configurable scaling limits |
-999.99 … +9999.99 engineering units |
||
Short circuit tolerant |
Yes |
||
Maximum external load |
500 Ω |
500 Ω (current output) |
|
Minimum external load |
0 Ω |
1000 Ω (voltage output) 0 Ω (current output) |
Humidity and temperature measurement |
See Probes > HC2A |