Technical specification

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Probes > HC2A-IC/IM/IE >

Technical specification

 

General


Device type

Humidity and temperature probe

Range of application

-100…+190 / 200 °C1,2 / 0…100 %rH

Mechanical configuration

See Dimensions

IP protection

IP65

Power supply / connection


Power supply (VDD)

3.3…5 VDC ±0 %

Current consumption

<4.5 mA

Maximum start-up current unlimited

<50mA during 2μs

Minimum start-up current

8 mA during 2ms and 5 mA during 2s

Maximum current spike during operation (steht so im Manual)

1 mA during maximum 2μs

Polarity protection

Mechanical protection

Humidity measurement


Sensor

ROTRONIC HYGROMER® HT-1

Measurement range

0…100 %rH

Measurement accuracy

±0.8 %rH at 10…30 °C

Repeatability

0.3 %rH

Long term stability

<1 %rh / year

Sensor response time

Typically 10 sec, for 63 % of 35 to 80 %rh step change (1 m/s wind speed at sensor)

Temperature measurement


Sensor

Pt100 RTD, IEC 751 1/3 Class B

Measurement range

-100…+190 / 200 °C1

Measurement accuracy

±0.1 °C at 10…30 °C

Repeatability

0.05 °C

Long term stability

<0.1 °C / year

Sensor response time

Typically 4 sec, for 63 %rH of a step change (1 m/s wind speed at sensor)

Calculated parameters


Psychrometric calculations

Dew or frost point (user configurable)

Start time / measurement interval


Start time

1.5 s (typical)

Measurement interval

1.0 s (typical) – when not calculating any parameter

Digital interface


Type

UART

Factory default

Baud rate 19200, tolerance 2 %

Parity: none

Data bits: 8

Stop bits: 1

Flow control: none

Baud rate configuration

No

Logic levels

Logic 0: ≤0.3 V * VDD

Logic 1: ≥0.8 V * VDD

Maximum cable length

5 m (16.4 ft) w/o signal booster

Conformity


CE / EMC

EMC Directive:

 

20.04.2016: 2014/30/EU

 

EN 61000-6-1: 2007, EN 61000-6-2: 2005

EN 61000-6-3: 2007+A1:2011+AC:2012, EN 61000-6-4: 2007+A1:2011

 

Soldering

Lead free (RoHS Directive 2011/65/EU)

FDA/GAMP directives

Compatible

Accepted environmental conditions


Storage/transportation

-50…+100 °C / 0…100 %rh, non-condensing

Range of application (electronics)

-50…+100 °C

0…100 %rh, non-condensing

Maximum wind velocity

See filter specification

Critical gases

Please consider the sensor data sheet of the HT-1 sensor.

1 The peak load at 200 °C is 100 h.

2 Only for HC2A-IEXXX: The sealing ring withstands only -40…100°C

 

 

Configurable analog outputs

Output 1

Can be made to correspond to any parameter


Factory default parameter

Relative humidity


Factory default scale

0…100 %RH

Output 2

Can be made to correspond to any parameter


Factory default parameter

Temperature


Factory default scale

-40…60 °C

Output 1 and Output 2



Signal type

0…1 V


User configurable scaling limits

-999.99 … 9999.99 engineering units


Offset at 0 V

+10 mV (maximum)


Ripple

<1.2 mV/x 2μs


Short circuit tolerant

Yes


Internal resistance

<10 Ω


Minimum external load

1000 Ω


Deviation from digital signal

<±1 mV from 0.002 to 1.0 V / + 2 mV from 0.0 to 0.002 V