Maintenance

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Transmitter > XB32A >

Maintenance

Service cable

 

IMPORTANT:

 

Use service cable AC3009 with all 2-wire, loop powered XB transmitters. This cable powers up the transmitter via the service connector. Do not use any other method for powering the transmitter when using this cable AC3009 as this will create a ground loop and damage the transmitter. For the same reasons do not use cable AC3006 with a 2-wire, loop powered transmitter.

Use service cable AC3006 with all 3-wire XB transmitter. This cable does not provide power to the transmitter and the transmitter should powered separately when using this cable

Both cables AC3006 and AC3009 convert UART (service connector) to USB and are used to connect the transmitter to a USB port of a PC running the ROTRONIC HygroSoft software.

 

 

 

Location of the service connector (mini USB type)

 

WARNING: the service connector is a UART interface with a mini-USB connector type. Do not connect the service connector directly to the USB port of a PC or hub.

 

XB-location-service-connector