General description

<< Click to Display Table of Contents >>

Navigation:  Water Activity Devices > HC2A-AW / HC2A-AW-USB >

General description

The HC2A-AW and HC2A-AW-USB is a digital humidity-temperature probe based on the AirChip 3000 technology.

 

HC2A-AW

HC2A-AW_Connector

The HC2A-AW is a water activity measurement head, which comes with UART (HC2-) connector and 1m cable.

It can be used with the HYGROLAB or in combination with an AC3001 cable also with the PC with HygroSoft software installed.

 

The HC2A-AW probe is directly powered from the HygroLab or by theUSB port of a PC (when using it with the AC3001 cable).

When connected to the HygroLab, the probe can be powered or unpowered by pressing on the touch button ON / OFF located on top of

the probe. The green LED ring is lit when the probe is powered.

 

 

 

HC2A-AW-USB

HC2A-AW-USB_Connector

The HC2A-AW-USB is a water activity measurement head, which comes with USB connector and 3m cable.

It can be only used with the PC and the HygroSoft. It's not compatible with the HygroLab.

 

Note: Please make sure that the USB driver is installed, see here: Driver Installation.

 

The HC2A-AW-USB probe is directly powered from the USB port of a PC. When connected to a

USB port, the probe can be powered or unpowered by pressing on the touch button ON / OFF located on top of

the probe. The green LED ring is lit when the probe is powered.

 

 

 

 

Sample holder and disposable sample cups

 

WP-40S sample holder with removable depth reduction insert.

 

WP-40

 

 

The PS-14 cups (shallow) are suitable for calibrating the probe with the Rotronic certified

humidity standards or with saturated salt solutions. In general, use the shallow cups for a

liquid, a paste or a powder. Use the PS-40 (deep) sample cups for products in bulk (large

chunks).

 