<< Click to Display Table of Contents >> Navigation: Humidity & Temperature Devices > Transmitter > HF3A > Electrical Connections |
This section provides detailed instructions on how to properly connect the HF3A device to power sources and other necessary components. It covers all possible wiring options. Proper attention to these steps is essential to guarantee reliable and safe functionality of the device.
There are 3 different transmitter variants available, each available with or without display and with 3 different mechanical installation:
Standard Variants / Order Code |
Description |
Mount |
Output |
HF3A-21D0300210100 |
2-wire with Display |
Wall |
Analog: 4...20 mA
|
HF3A-21X0000210100 |
2-wire without Display |
||
HF3A-22D0300210100 |
2-wire with Display |
Cable (2m) |
|
HF3A-22X0000210100 |
2-wire without Display |
||
HF3A-23D0300210100 |
2-wire with Display |
Duct |
|
HF3A-23X0000210100 |
2-wire without Display |
||
HF3A-31D0300210100 |
3-/4-wire with Display |
Wall |
Analog: 0...20, 4...20 mA, 0...1, 0...5, 0...10 V |
HF3A-31X0000210100 |
3-/4-wire without Display |
||
HF3A-32D0300210100 |
3-/4-wire with Display |
Cable (2m) |
|
HF3A-32X0000210100 |
3-/4-wire without Display |
||
HF3A-33D0300210100 |
3-/4-wire with Display |
Duct |
|
HF3A-33X0000210100 |
3-/4-wire without Display |
||
HF3A-D1D0300210100 |
3-/4-wire with Display |
Wall |
Analog: 0...20, 4...20 mA, 0...1, 0...5, 0...10 V
Digital: Modbus RTU |
HF3A-D1X0000210100 |
3-/4-wire without Display |
||
HF3A-D2D0300210100 |
3-/4-wire with Display |
Cable (2m) |
|
HF3A-D2X0000210100 |
3-/4-wire without Display |
||
HF3A-D3D0300210100 |
3-/4-wire with Display |
Duct |
|
HF3A-D3X0000210100 |
3-/4-wire without Display |
Terminal Overview |
Analog Terminal Description |
|
![]() |
V1+
OUT1
V2+
OUT2 |
Output Humidity +
Output Humidity -
Output Temperature +
Output Temperature - |
Connection diagram |
||
Two separated power supply |
One power supply |
|
![]() |
![]() |
Note: The service cable AC3009 must not be connected simultaneously with the power supply.
Terminal Overview |
Analog Terminal Description |
|
![]() |
V+
V-
OUT1+
OUT1-
OUT2+
OUT2- |
Power Supply +
Power Supply -
Analog output 1+
Analog output 1-
Analog output 2+
Analog output 2- |
Connection diagram |
||
Current |
Voltage |
|
![]() |
![]() |
Note: The service cable AC3009 must be used in case that the power supply is simultaneously connected.
Terminal Overview |
Analog Terminal Description |
|
![]() |
V+
V-
OUT1+
OUT1-
OUT2+
OUT2- |
Power Supply +
Power Supply -
Analog output 1+
Analog output 1-
Analog output 2+
Analog output 2- |
Digital Terminal Description |
||
RXTX+
RXTX-
S-GND |
RS485+ (A)
RS485- (B)
Signal Ground (-) |
|
Connection diagram |
||
Current |
Voltage |
|
![]() |
![]() |
Note: The service cable AC3009 must be used in case that the power supply is simultaneously connected.