Device Description

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Handheld & Logger > HP31 >

Device Description

The HF3 series offers the following user functions:

 

• User configurable settings

• Calculation of the dew or frost point

• Humidity temperature calibration and adjustment

• Simulator mode

 

 

Variants

Two types of electronic circuit are available: HF32: 2-wire, loop powered (4…20 mA current signal) and HF33: 3-wire (voltage or current signal). Both circuit types provide linear analog outputs signals for transmission over a length of cable to a remote display, recorder, controller or data processing unit. The HF3 transmitters can be used to measure humidity only, temperature only or both parameters.

 


Wire

Mount type

Output

Display

HF320

2-Wire

S/R/W/D

4...20mA

only for S & R type, no backlight

HF332

3-Wire

S/R/W/D

4...20mA

only for S & R type, with backlight

HF335

3-Wire

S/R/W/D

0...10V

only for S & R type, with backlight

S = Space, R = Room, W = Wall, D = Duct

 

 

Display option

HF3-display-option

 

The LC display option is available only with type R and type S mechanical configurations. The upper line of the display corresponds to relative humidity or dew / frost point and the bottom line corresponds to temperature.

 

The display can be configured to show a trend indicator on each line:

 

▲: increasing value

▼: decreasing value

 

In the event of an alarm the display shows the symbol [ ! ] to the right of the value.

For instructions see the following HW5 or HygroSoft manual.

 

The display option for the HF33 has a backlight. The display option for the HF32 does not have a backlight.

 

 

 

Power supply

HF32

2-wire, loop powered


10…28 VDC - depending on the load connected to the output(s).

The minimum supply voltage can be determined as follows:

 

V min = 10 V + (0.02 x Load*) *Load resistance in Ω.

 

For the maximum load of 500 Ω, the minimum supply voltage is 10 + (0.02 x 500) = 20 VDC. With both output circuits closed, the maximum current consumption is 40 mA.

 

HF33

3-wire, analog outputs


18 to 40 VDC (see note below) or 13 to 28 VAC. With both output circuits closed, the maximum current consumption is 60 mA for DC powered devices and 150 mA for AC powered devices. HF3 with display and backlight have a current consumption up to 100 mA (DC powered) / 250 mA (AC powered).

 

Note: depending on the type of output signal, the HF33 will operate with the following minimum voltage

 

0…1 V outputs:

6 VDC or 5 VAC

0…5 V outputs:

10 VDC or 8 VAC

0…10 V outputs:

18 VDC or 13 VAC

0…20 mA or 4 …20 mA outputs:

6 VDC or 5 VAC with 0 Ω load


18 VDC or 13 VAC with 500 Ω load

 

 

 

 

Measured parameters

The HF3 measures relative humidity with a ROTRONIC Hygromer® IN-1 capacitive sensor and temperature with a Pt100 RTD.

 

 

 

Calculated parameters

Using the ROTRONIC HW5 or Hygrosoft software, the HF3 can be configured by the user to calculate either the dew point or the frost point.

 

 

 

Analog output signals (HF32 and HF33)

HF32 and HF33:

With the ROTRONIC HW5 or Hygrosoft software any of the two analog output signals can be made to correspond to one of the following:

 

• Relative humidity

• Temperature

• Dew or frost point

 

Any output can also be disabled.

 

The scale of each analog output can be set within the numerical limits of -999.99 and 9999.99.

 

The D/A converters used to generate the analog output signals feature a 16-bit resolution and exhibit a small positive offset at the bottom of the signal range as indicated below:

 

 

Signal type

Maximum offset at range bottom

0…1 V

10 mV

0…5 V

50 mV

0…10 V

100 mV

0…20 mA

0.2 mA

4…20 mA

No offset

 

HF33:

The ROTRONIC HW5 or HygroSoft software allows changing the type of output signal to one of the following: 0…20 mA, 4…20 mA, 0…1V, 0…5V or 0…10V. Both output signals are automatically configured with the same signal type. No calibration or adjustment is required after changing the type of output signal.

 

In the case of voltage output signals, load requirements apply to the external device or circuit connected to the HF3 transmitter. These requirements are defined in the “ operating modes” chapter.

 

 

 

Service connector

The service connector is a UART digital interface (Universal Asynchronous Receiver Transmitter) with a mini-USB type connector. This allows connecting the HF3 either to a PC running the ROTRONIC HW5 or HygroSoft software or to a probe input of another instrument that is compatible with the HygroClip Advanced (HC2A) probes. In both cases a service cable is required. See “ Maintenance” for the location of the service connector and for the type of service cable to be used.

 

Connecting the HF3 to a PC is used to configure the HF3, gain access to the HF3 functions such as humidity and temperature adjustment, read data from the HF3 on the PC and update the AirChip3000 firmware.

Connecting the HF3 to another instrument is useful only when the other instrument has its own display and keypad. The connection allows showing the data measured by the HF3 on the other instrument display and also allows using the other instrument internal menu to do for example a humidity and temperature adjustment of the HF3.

 

 

Sensor protection (dust filter)