Default Configuration

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Transmitter > HF3 > Customer adjustment >

Default Configuration

Configuration of the HF3 by the user and access to its functions requires a PC with the ROTRONIC HW5 or HygroSoft installed. Service cable AC3006 or AC3009 is used to connect the HF3 service connector to a USB port of the PC (see Maintenance -> Service Cable).

 

Configurable Settings

Applicability

Factory default

Unit system (Metric or English)

HF32, HF33

As per ordering code

Analog signal type (4…20 mA or other)

HF33

As per ordering code

Psychrometric calculation

HF32, HF33

As per ordering code

Output 1 parameter, scale and unit

HF32, HF33

As per ordering code (%RH or DP)

Output 2 parameter, scale and unit

HF32, HF33

Temperature, unit as per ordering code

Display resolution (optional)

HF32-S, HF33-S

1 decimal

Display backlight (optional)

HF33-S

Always on

Displayed parameters

HF32, HF33

As per ordering code

Trend indicator (display)

HF32-S, HF33-S

Enabled

Device name

HF32, HF33

Instrument model

 

 

 

Functions

Applicability

Factory default

Humidity / temperature adjustment

HF32, HF33


Device write protection

HF32, HF33

Disabled

Limit humidity output to 100 %RH

HF32, HF33

Enabled

Out-of-limit value digital / display alarm

HF32-S, HF33-S

Disabled

Simulator mode

HF32, HF33

Disabled

 

 

For a detailed description of all AirChip3000

The factory default setting for dew / frost point calculation is frost point below freezing