Contamination/Pollutants

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > Sensors >

Contamination/Pollutants

Some gases and contaminants/pollutants can damage Rotronic humidity sensors.

The contaminants/pollutants can be divided into two categories:

 

Gases without influence and gases with an influence on the humidity sensors.

 

For contaminants/pollutants with an influence on the sensors and therefore with an influence

on the measurement result, the maximum constant concentration must be known (see table below).

 

Contaminants/Pollutants with an influence

 

Contaminant/Pollutant

Formula

MAC value

Permissible constant concentration

IN-1

 

HH-1

HT-1

ppm

mg/m3

ppm

 

mg/m3

ppm

mg/m3

ppm

mg/m3

Acetone

CH3COCH3

1000

2400

3300

8000

3700

9000

3300

8000

Ammonia

NH3

25

18

5500

4000

5500

4000

5500

4000

Petrol

 

300

1200

 

150000

 

150000

 

150000

Chlorine

Cl2

0.5

1.5

0.7

2

1.5

4.5

0.7

2

Acetic acid

CH3COOH

10

25

800

2000

1000

2500

800

2000

Ethyl acetate

CH3COOC2H5

400

1400

4000

15000

4000

15000

4000

15000

Ethanol

C2H5OH

1000

1900

3500

6000

5800

10000

3500

6000

Ethylene glycol

HOCH2CH2OH

100

260

1200

3000

1200

3000

1200

3000

Formaldehyde

HCHO

1

1.2

2400

3000

2400

3000

2400

3000

Isopropanol

(CH3)2CHOH

400

980

4800

12000

6000

15000

4800

12000

Methanol

CH3OH

200

260

3500

6000

6000

8000

3500

6000

Methyl ethyl keton

C2H5COCH3

200

590

3300

8000

3300

8000

3300

8000

Ozone

O3

0.1

0.2

1

2

1.5

3

1

2

Hydrochloric acid

HCl

5

7

300

500

300

500

300

500

Sulfur dioxide

SO2

5

13

5

13

5

13

5

13

Hydrogen sulfide

H2S

10

15

350

500

350

500

350

500

Nitrous gases

NOx

5

9

5

9

5

9

5

9

Toluene

C6H5CH3

100

380

1300

5000

1800

7000

1300

5000

Hydrogen peroxide

H2O2

1

1.4

90

130

880

1200

90

130

Xylene

C6H5(CH3)2

100

440

1300

5000

1800

7000

1300

5000

 

Contaminants/Pollutants without influence

 

Note that the common sealing material silicone damages the sensor! When probes are installed

silicone must not be used!

 

Contaminant/Pollutant

Formula

Argon

Ar

Butane

C4H10

Natural gas

 

Ethane

C2H6

Helium

He

Methane

CH4

Neon

Ne

Propane

C3H8

Oxygen

O2

Nitrogen

N2

Hydrogen

H2