<< Click to Display Table of Contents >> Navigation: Water Activity Devices > AwTherm > Specification |
General |
|
Device type |
Temperature-stabilized water-activity measurement |
Operating conditions |
1..40 °C (34..107 °F) |
Measurement range |
0.005…1.000 aw |
Accuracy |
±0.005 aw ±0.1 °C (±0.18 °F) |
Power supply |
110..230 V / 50..60 Hz |
Display |
8-line LCD with touch operation |
Sample sizes |
Variable (14 mm / 40 mm) |
Technical Information |
|
Current Consumption |
≤2 A |
Temperature-control range |
0..60 °C (32..140 °F) |
Temperature stability |
±0.01 °C/min (±0.018 °F/min) |
Chamber-temperature gradient |
<0.1 °C (<0.18 °F) |
Firmware update |
Via USB port |
Functions |
|
HygroSoft compatible |
Yes |
aw-Quick function |
Yes |
Interface |
Micro USB |
Calibration / Adjustment |
Aw & Temperature: HG2-S (probe removed with AwT-CAL) Aw: ROTRONIC humidity standards (via HygroSoft) |
Trend indicator |
Yes |
Probes |
|
Sensor |
HYGROMER® IN-1 |
Maintenance / Calibration |
Annual calibration (recommended) |
Long-Term Stability |
<0.01 aw/year |
Temperature sensor |
Pt100, DIN 1/3 Class B |
Approval / Conformity |
|
Standards |
ISO 18787 |
CE / EMC |
EMC 2014/30/EU RoHS 2011/65/EU |
IP protection |
IP21 |
Housing / Mechanical parts |
|
Housing material |
PC / ABS |
Housing dimensions |
400 x 180 x 180 mm |
Sample container holder dimensions |
AwT-PS14: Ø46 x 14 mm AwT-PS40: Ø46 x 40 mm |
Weight |
4200g |
This standard applies to equipment designed to be safe at least under the following conditions:
•For indoor use only
•Altitude up to 2000 m
•Ambient temperature 5°C to 40°C
•Maximum relative humidity 80 % for temperatures up to 31 °C decreasing linearly to 50 % Relative Humidity at 40 °C
•Pollution degree 2
• Input voltage is from 90VAC to 264VAC; Auto Ranging
• Input Frequency range from 47Hz to 63Hz
• Input current is 3.5A at 100VAC or 1.5A at 240VAC
• Power inlet IEC socket
Batteries must not be disposed of as ordinary refuse. Please use a local battery recycler.
Replacement batteries are available from Rotronic. These are Lithium rechargeable batteries and must only be replaced with the same type. The replacement battery will be supplied with plastic tweezers; do not use a metal tool to remove the battery as this will cause a short circuit. Be sure to insert the new battery using the correct polarity as shown.
Use of a non-rechargeable battery, or incorrectly inserting the battery could lead to explosion.