ATEX Data

<< Click to Display Table of Contents >>

Navigation:  Humidity & Temperature Devices > ATEX Devices >

ATEX Data

 

Ignition Protection Category, Type

 

Measurement probe with HygroClip:        For Zone 0/20

Transmitter:                                        Complete transmitter for Zone 1/2 and 21/22

 

The following ATEX approval according to IEC is fulfilled:

 

Device group II for explosive atmospheres (not mining)

 

Probe: For application temperature Tamb = [-40 … +60 °C]

HF5EX_Ex_Logo II 1/2 G Ex ia IIC T5 Ga/Gb                        Zone 0, gas, intrinsically safe, temp. 100°C

HF5EX_Data II 1/2 D Ex ia IIIC T80°C Da/Db                Zone 20, dust, intrinsically safe, temp. 80°C

 P66                                                        IP protection 66

 

Or

 

For application temperature Tamb = [-40 … +85 °C]

HF5EX_Ex_Logo II 1/2 G Ex ia IIC T4 Ga/Gb                        Zone 0, gas, intrinsically safe, temp. 130°C

HF5EX_Ex_Logo II 1/2 D Ex ia IIIC T110°C Da/Db                Zone 20, dust, intrinsically safe, temp. 110 °C

 IP66                                                        IP protection 66

 

 

(1/2 - 1: Zone 0 and 2: Zone 1 suitable for mounting in zone separation wall)

(Ga - very high protection level (Zone 0), Gb - high protection level (Zone 1))

(Da - very high protection level (Zone 0), Db - high protection level (Zone 1))

(Ga/Gb, Da/Db suitable for mounting in zone separation wall)

 

Transmitter:        For application tempearture Tamb = [-40 … +60 °C] (without LCD) and

                 For application temperature Tamb = [-10 … +60 °C] (with LCD)

 

HF5EX_Ex_Logo II 2(1) G Ex eb mb [ia Ga] IIC T5 Gb                Zone 1, 2, gas, (intrinsically safe), temp. 100°C

HF5EX_Ex_Logo II 2(1) D Ex tb [ia Ga] IIIC T80°C Db                Zone 21, 22, dust, (intrinsically safe), temp.80°C

 IP66                                                        IP protection 66

 

(2(1) - 2: Zone 1, (1): contains electric circuits that may exist in Zone 0)

(Ex e mb [ia Ga] multiple ignition protection categories: Ex-e, Ex-mb and output Ex-ia)

 

 

Complete system:        HF5EX_Ex_Logo II 1/2 G Ex eb ia mb IIC T5 Ga/Gb

                         HF5EX_Ex_Logo II 1/2 D Ex ia tb IIIC T80°C Da/Db

 

 

 

ATEX-Relevant Standards

The following European standards are relevant for the product:

CENELEC - European Committee for Electrotechnical Standardization

 

EN 60079–0                Part 0:                Equipment – General requirements Construction, testing, marking

 

EN 60079–7                Part 7:                Equipment protection by increased safety “e”

 

EN 60079–11                Part 11:                Equipment protection by intrinsic safety “i”

 

EN 60079–14                Part 14:                Electrical installations design, selection and erection

 

EN 60079–18                Part 18:                Equipment protection by encapsulation “m”

 

EN 60079–26                Part 26:                Equipment with equipment protection level (EPL) Ga Zone 0 requirements (EPL - Equipment protection level)

 

EN 60079–31                Part 31:                Equipment dust ignition protection by enclosure “t”

 

 

Nominal Data

Power supply:

There are two separate connections available for the humidity and temperature sensors. The device connections comprise power supply and sensor signal. The supply voltages for the two sensors may be different because they are isolated by two internal diodes. The device can also be operated with only one loop.

 

Device input voltage:                USupply                = 20 V … 28 VDC [24 V +/- 15%]

 

Maximum current:                ISupply max.                = 50 mA [Sum of both input currents]

 

Load impedance:                          RM                        = 0…500 Ω [Measurement voltages 0 V…10 V]

 

Ambient temperature range:

Measurement probe:                TA = [- 40 °C ... + 85 °C]

Transmitter with LCD:                TA = [- 10 °C … + 60 °C]

Transmitter without LCD:        TA = [- 40 °C … + 60 °C]

 

The normal ATEX standards only apply for normal temperatures [-20 °C … +40 °C] and for a pressure range of [0.8 bar…1.1 bar]. The extended temperature range must be marked on the product identification label.

 

Input protection circuit:

The device is protected against polarity reversal by two diodes. In addition to this, two Z-Transil

diodes protect against static and transient overvoltages. These measures are not required by ATEX.

They were, however, implemented as easily realizable additional protective measures.

 

Um = 28 V DC

Supply circuit sensor*

HC2-LDPxxx-EX

Sensor circuit*

Maximum output voltage:

Uo ≤ 6.6 V

Maximum input voltage:

Ui ≤ 6.6 V

Maximum output current:

Io ≤ 82 mA

Maximum input current:

Ii ≤ 82 mA

Maximum output power:

Po ≤ 485 mW

Maximum input power:

Pi ≤ 485 mW

Capacity:

Co = 22 μF

Interne Kapazität:

Ci = 19 μF

Inductance:

Lo = 2 μH

Internal inductance:

Li = 0





All other sensors

Sensor circuit*



Maximum input voltage:

Ui ≤ 6.6 V



Maximum input current:

Ii ≤ 82 mA



Maximum input power:

Pi ≤ 485 mW



Internal capacity:

Ci = 15 μF



Internal inductance:

Li = 0

* Type of protection intrinsic safety Ex ia IIC.

 

 

 

Product Identification Label

 

Manufacturer:                                        Rotronic AG, CH-8303 Bassersdorf

Type designation:                                        HF520-EX

Serial number:                                        < Number >

 

Measurement probe:                                HF5EX_Ex_Logo II 1/2 G Ex ia IIC T5…T4 Ga/Gb

                                                 HF5EX_Ex_Logo II 1/2 D Ex ia IIIC T80 °C…T110 °C Da/Db

                                                               IP66

 

Transmitter:                                        HF5EX_Ex_Logo II 2(1) G Ex eb mb [ia Ga] IIC T5 Gb

                                                 HF5EX_Ex_Logo II 2(1) D Ex tb [ia Ga] IIIC T80°C Db

                                                               IP66

 

Certification body and certificate no.:        Eurofins, Fehraltorf (CH)

                                                 SEV 14 ATEX 0107

                                                 IECEx SEV 14.0002

                                                 CML 22UKEX3190

 

Date:                                                        09.09.2024

 

Operating temperature range:                Tamb = [-10 °C … +60 °C] with LCD

                                                 Tamb = [-40 °C … +60 °C] without LCD

 

In:                                                        [20 VDC … 28 VDC], 2 W

Out:                                                        [4 mA … 20 mA], 2-wire current loop

 

Symbols:                                                HF5EX_UKCA2503                HF5EX_CE