V2.3.1.22325

<< Click to Display Table of Contents >>

Navigation:  Software Overview > RMS Monitoring Software > Software Release Notes >

V2.3.1.22325

MS6 release date: 23.11.2022.

 

Improvements

ID

Description

RMS-3248

Reset a password without an E-Mail server.

RMS-3075

Firmware files not deleted when the software version is updated (cloud only).

RMS-3063

Pentest: Input Validation Missing.

RMS-3062

Pentest: Inconsistent Authorization.

RMS-3048

Option to send E-Mail's over Port 587.

RMS-2983

The MPT/DEV selection dialogue boxes are too small.

RMS-2639

Work needed for every release.

 

Bug Fixes

ID

Description

RMS-3294

Phone notification continues when alarm is inhibited from the table page.

RMS-3247

Redundant server: the service starts in passive mode if the database is not available. When the database is then available the service switches to active mode and an error occurs.

RMS-3149

When a user is deactivated/deleted, the notification settings remain.

RMS-3148

Reports created by a user are deleted when the user who created the report is deleted.

RMS-3147

Automated reports are not being created based upon the exact time frame requested.

RMS-3122

Retroactive alarms are not triggered when values go directly from high to low alarm or vice versa.

RMS-3092

Issues with wireless loggers when added to another cloud account (See RMS-2957).

RMS-3049

Event «Update to V2.2.0» is repeated every 5 seconds.

RMS-3034

The deviation report wrong colors for alarm zone.

RMS-3019

No finished E-mail at the end of an alarm via alarm schema’s.

RMS-2989

When a negative alarm is set and the event is over, the deviation report will remain incorrect (incorrect measured value, setpoint and limit deviations).

RMS-2988

There is an error in the deviation report when the event is not closed. The date is set to 01.01.0001.

RMS-2957

Unlike with the LAN loggers there is no warning when wireless loggers are added to another Cloud Account.

RMS-2935

Relay not closing when the alarm event is closed.

RMS-2882

When removing the ADC input 1, the input 2 shifts to input 1.

RMS-2523

No audit trail when the report template and logo’s are changed.