V2.0.0.19353

<< Click to Display Table of Contents >>

Navigation:  Software Overview > RMS Monitoring Software > Software Release Notes >

V2.0.0.19353

MS6 release date: 30.09.2019.

 

New Features

ID

Description

RMS-2259

New software architecture.

 
Details: the RMS software structure has been improved. The main benefits are based upon the scaling for the cloud:

The system is improved for larger amounts of devices and measuring points.

Stability improvement.

 

Improvements

ID

Description

RMS-2533

Change gateway concept:

The average of the last 5 RSSI values is taken.

If one value is missing, the average RSSI value will be reduced by 10.

The gateway will be switched if the difference is >=15dB (or 10dBm if the RSSI <-90dBm).

RMS-2532

Adapted the download commands:

The download commands will be sent over a period of 3 days. After 3 days a data gap will be created.

The download command will be sent at increasing intervals. The intervals will be doubled until a maximum of 2 hours (ex: 1, 2, 4, 8, 16, 32, 64, 120, 120, 120...).

Within 3 days, the around 40 download trials will happen.

RMS-2534

The RMS-Monitor has been removed from the system.

RMS-2586

The RMS software will work with the latest Google Chrome update (Javascript).

 

Bug Fixes

ID

Description

RMS-2389

Typo: "AES Key Refresh" not "AES Key Refeseh".

RMS-2383

Unit missing in PDF reports.

RMS-2294

Property definitions cannot be overwritten.

RMS-2293

Firmware update: "NoPropertyData".

RMS-2234

Not possible to archive a measuring point.

RMS-1178

MPT-number not available with an SMS.

RMS-1117

Data from an unassigned gateway will not be taken (if encrypted).

RMS-1093

Archiving crashed when copying measuring data.

RMS-1051

Software exception: UpdateCalculations.