General |
|||
---|---|---|---|
Device type |
RMS digital input module |
||
Range of application |
-40…70 °C / 0…100 %rh, Non-condensing |
||
IP protection |
IP20 |
||
Storage and transport conditions |
-40…30 °C / 0…90 %rh |
||
Maximum altitude of operation |
2000 m.a.s.l. |
||
Interfaces |
Ethernet |
||
Data memory |
75‘000 measured values |
||
Number of inputs |
2 independent digital inputs |
||
Input frequency |
Max: 0.833 Hz, btw. 1.2 s |
||
Pulse detection time |
Min: 100 ms |
||
Input circuit |
Logic levels: 0 V / 5-24 V Triggering-threshold: ~3.77 V Current consumption: <1 mA |
||
Reed circuit |
Maximum load at input: 100 kΩ |
||
Power supply |
|||
Supply voltage |
24 VDC ±10 % / <100 mA1 PoE: 802.3af-2003, Class 1 Battery |
||
Battery life |
The batteries of RMS devices typically last 3 years under the optimal conditions: •Temperature = 20°C±5°C. •Wireless signal strength >-60 dBm. •Limited “read out data” events in the audit trail. |
||
Current consumption |
<100 mA |
||
Polarity protection |
Yes |
||
Start time / measurement interval |
|||
Startup time |
10 s (typical) |
||
Logging frequency |
Event triggered & interval (10 s to 15 min) |
||
General specifications |
|||
Housing material |
PC (Polycarbonate) |
||
Dimensions |
108 x 89.7 x 62.2 mm |
||
Weight |
200 g |
||
Fire protection class |
UL94-V0 |
||
Maximum input cable length |
<3 m min. Cat. 5 |
||
Conformity |
|||
EMC directives: 2014/30/EU
LVD- directives: 2014/35/EU |
Performance criterion: www.rotronic.com |
||
Soldering material |
Lead free (RoHS Directive 2014/65/EU) |
||
FDA/GAMP directives |
FDA CFR 21 Part 11 / GAMP5 |
1 Power supply requirements: 24 VDC ±10 % / >4 W / limited power source