Technical specifications

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Input and Output Devices > RMS-DI-L-R >

Technical specifications

 

General


Device type

RMS digital input module

Range of application

-40…70 °C / 0…100 %rh, Non-condensing

IP protection

IP20

Storage and transport conditions

-40…30 °C / 0…90 %rh

Maximum altitude of operation

2000 m.a.s.l.

Interfaces        

Ethernet

Data memory        

75‘000 measured values        

Number of inputs        

2 independent digital inputs

Input frequency

Max: 0.833 Hz, btw. 1.2 s

Pulse detection time

Min: 100 ms

Input circuit

Logic levels: 0 V / 5-24 V

Triggering-threshold: ~3.77 V

Current consumption: <1 mA

Reed circuit

Maximum load at input: 100 kΩ

Power supply


Supply voltage

24 VDC ±10 % / <100 mA1

PoE: 802.3af-2003, Class 1

Battery

Battery life

The batteries of RMS devices typically last 3 years under the optimal conditions:

Temperature = 20°C±5°C.

Wireless signal strength >-60 dBm.

Limited “read out data” events in the audit trail.

Current consumption

<100 mA

Polarity protection

Yes

Start time / measurement interval


Startup time

10 s (typical)

Logging frequency

Event triggered & interval (10 s to 15 min)

General specifications


Housing material

PC (Polycarbonate)

Dimensions

108 x 89.7 x 62.2 mm

Weight

200 g

Fire protection class

UL94-V0

Maximum input cable length

<3 m min. Cat. 5

Conformity


EMC directives: 2014/30/EU

 

LVD- directives: 2014/35/EU

EN 61326-1        

EN 61000-6-2

EN 55011        

EN 55032        

EN 61010-1

IEC 61326-1

IEC 61000-6-2

IEC CISPR 11

IEC CISPR 32

IEC 61010-1

 

Performance criterion: www.rotronic.com

Soldering material

Lead free (RoHS Directive 2014/65/EU)

FDA/GAMP directives

FDA CFR 21 Part 11 / GAMP5

1 Power supply requirements: 24 VDC ±10 % / >4 W / limited power source