<< Click to Display Table of Contents >> Navigation: Hardware Overview > Input and Output Devices > RMS-DO-L-R > Technical specifications |
General |
|||
---|---|---|---|
Device type |
RMS digital output module |
||
Range of application |
-40…70 °C / 0…100 %rh, Non-condensing |
||
IP protection |
IP20 |
||
Storage and transport conditions |
-40…30 °C / 0…90 %rh |
||
Maximum altitude of operation |
2000 m.a.s.l. |
||
Interfaces |
Ethernet |
||
Protocols |
Modbus |
||
Number of outputs |
2 |
||
Relay switching power |
50 VAC(Peak) 1 A / 50 VDC/1 A, polarity independent |
||
Voltage output (VEX) |
24 VDC (maximum current depends on the connected power supply) |
||
Power supply |
|
||
Supply voltage |
24 VDC ±10 % / <100 mA PoE: 802.3af-2003, Class 1 |
||
AC Adapter requirements |
24 VDC ±10 % / 4 W nominal /<15 W power-limited |
||
Current consumption |
<100 mA |
||
Polarity protection |
Yes |
||
Start time / measurement interval |
|||
Startup time |
10 s (typical) |
||
Update frequency |
Interval (10 s to 15 min) |
||
General specifications |
|||
Housing material |
PC (Polycarbonate) |
||
Dimensions |
108 x 89.7 x 62.2 mm |
||
Weight |
155 g |
||
Fire protection class |
UL94-V0 |
||
Maximum output cable length |
<3 m min. Cat. 5 |
||
Conformity |
|||
EMC directives: 2014/30/EU
LVD- directives: 2014/35/EU |
Performance criterion: www.rotronic.com |
||
Soldering material |
Lead free (RoHS Directive 2014/65/EU) |
||
FDA/GAMP directives |
FDA CFR 21 Part 11 / GAMP5 |