Technical Specification

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > Input and Output Devices > RMS-ADC-L-R >

Technical Specification

Specification overview



Sensor

1 or 2 analog inputs

Measurement range

0-1 V, 0-5 V, 0-10 V, 0-20 mA, 4-20 mA

Load

voltage input: 25 kOhm

current input: 250 Ohm

Measurement accuracy

+/- 0.03 % full-scale

Temperature accuracy

+/- 0.02 % full-scale /°C  

Sensor quality

oDetection if no sensor connected (open loop)

oDetection of overload on input

oDetection of under load on 4 to 20 mA configuration

Measurement interval

10 s ... 300 s

Power supply for sensors

24 V / 80 mA maximum

Power supply

24 VDC ±10 %

Current consumption

< 160 mA

External device power supply advised

24 VDC ±10 %, 4 W minimal, > 5 W Limited Power Source

Interface

10Base-T / 100Base-TX Ethernet

PoE

802.3af-2003, Class 1

Log memory

44'000 of measurement for each channel

Startup time logger

< 20 s (fixed IP address)

< 40 s (DHCP)

Input simulator function for validation

Fix values for each channel

Operational temperature logger

-40 ... 70 °C

Storage temperature

-40 ... 85 °C

Housing Material

PC-ABS

IP protection

IP20

Device weight

125 g

Dimensions

110 x 119 x 22.5 mm