<< Click to Display Table of Contents >> Navigation: Hardware Overview > Input and Output Devices > RMS-ADC-L-R > Technical Specification |
Sensor |
1 or 2 analog inputs |
Measurement range |
0-1 V, 0-5 V, 0-10 V, 0-20 mA, 4-20 mA |
Load |
voltage input: 25 kOhm current input: 250 Ohm |
Measurement accuracy |
+/- 0.03 % full-scale |
Temperature accuracy |
+/- 0.02 % full-scale /°C |
Sensor quality |
oDetection if no sensor connected (open loop) oDetection of overload on input oDetection of under load on 4 to 20 mA configuration |
Measurement interval |
10 s ... 300 s |
Power supply for sensors |
24 V / 80 mA maximum |
Power supply |
24 VDC ±10 % |
Current consumption |
< 160 mA |
External device power supply advised |
24 VDC ±10 %, 4 W minimal, > 5 W Limited Power Source |
Interface |
10Base-T / 100Base-TX Ethernet |
PoE |
802.3af-2003, Class 1 |
Log memory |
44'000 of measurement for each channel |
Startup time logger |
< 20 s (fixed IP address) < 40 s (DHCP) |
Input simulator function for validation |
Fix values for each channel |
Operational temperature logger |
-40 ... 70 °C |
Storage temperature |
-40 ... 85 °C |
Housing Material |
PC-ABS |
IP protection |
IP20 |
Device weight |
125 g |
Dimensions |
110 x 119 x 22.5 mm |