Function Overview

<< Click to Display Table of Contents >>

Navigation:  Hardware Overview > LOG Dataloggers > RMS-LOG-L > Customer adjustment >

Function Overview

Overview of the main software functions of the device

 

Discovery

With Discovery it is possible to find devices in the subnet with the RMS-Configuration software irrespective of their IP configuration and to change their settings.

IP configuration        

 

The devices can have static or dynamic IP configurations. It is recommended that you use a dynamic IP configuration whenever possible. If fixed IPs are used, the network topology must be considered exactly.

RMS Web Server settings        

 

Every device has the server address and software path of the RMS server software stored in it in order to build up communication with the RMS server software.

 

The two parameters can be set with the RMS-Configuration software:

oHost: Address of the server with the RMS software

oServer path: Server path where the server software is installed.

Measurement by the connected probe        

The device recognizes the connected probe automatically and requests the latest measured data. The measured data are sent to the RMS server software at the fixed interval directly after the measurement.

Save measured data        

 

The measured values of every measurement are saved in the internal ring memory (44,000 pairs of measured values). If the data cannot be sent to the server software directly, they are kept in the device and then sent later as soon as the connection to the server software has been restored.

Battery mode        

 

If the external power supply fails (24 VDC / PoE), the device runs in battery mode. Measurements are still carried out at the fixed interval and the data saved in the ring memory (44,000 pairs of measured values).

Firmware update